78 . RamanFlex 400 Series Getting Started Guide
Calibration and Performance Verification SOPs
Several different automated calibration tests are included in the software, these are:
•
Wavelength (x-axis) calibration uses the position of precisely known lines in a neon
emission spectrum to calibrate the detector.
•
Intensity (y-axis) calibration is performed using specially doped glass with a well
characterized emission spectrum. This is used to correct for non-linearity in detector
response.
•
Laser frequency calibration determines the exact laser wavelength.
An additional wavelength calibration verification routine has been included in the software,
which compares a series of peaks for polystyrene (ASTM E 1840) with values collected using
a polystyrene sample. These calibration/validation tests allow for in-house instrument
performance verification.
Intensity calibration is only performed if the instrument has been moved or as part of an
annual service, or if there have been appreciable temperature variations in the local
environment.
Backgrounds are always collected during a calibration procedure. The background measures
the bias and dark signal over the selected area of the CCD. This is a necessary part of the
calibration routines, so it is always performed.
The spectrometer has no moving parts in its detection system, therefore frequent calibration
is not necessary.
We recommend that you calibrate the instrument in accordance with the following SOP.
SOP-7: Calibration/Performance Verification
We recommend that wavelength calibration verification is performed on a regular basis to
ensure that the instrument is performing within specification. This can be once a day or once
a week depending on your requirements and the laboratory environment.
On completion of the wavelength calibration verification, a calibration verification report will
provide details on the outcome of the verification procedure.
IMPORTANT: If the instrument passes the wavelength calibration verification procedure, the
instrument is optimized for performance and does not require wavelength calibration.
Содержание RamanFlex 400
Страница 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 27: ...Raman History Theory and Instrumentation ...
Страница 34: ...34 RamanFlex 400 Series Getting Started Guide ...
Страница 35: ...Preparing your Facility ...
Страница 40: ...40 RamanFlex 400 Series Getting Started Guide ...
Страница 41: ...Installing your RamanFlex 400 ...
Страница 61: ...Introduction to the RamanFlex 400 ...
Страница 68: ...68 RamanFlex 400 Series Getting Started Guide ...
Страница 69: ...Using the RamanFlex 400 ...
Страница 85: ...Maintenance and Troubleshooting ...