Using the RamanFlex 400 . 79
Figure 18 System verification / calibration flow diagram
If wavelength calibration verification fails and there is an error message stating that peak
wavelengths are outside limits, then wavelength and laser frequency calibrations should be
performed. When these calibrations have finished, repeat the wavelength calibration
verification to ensure that the instrument has been brought back within specification.
If the wavelength calibration verification fails because the spectral intensity is low, then this
may also be due to the calibration sample holder not being correctly positioned, optimally
focused or because the laser key is not switched on. Then repeat the wavelength calibration
verification to ensure that the instrument has been brought back within specification. If the
instrument fails the wavelength calibration verification again, please contact PerkinElmer.
Содержание RamanFlex 400
Страница 1: ...RamanFlex 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanFlex 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 27: ...Raman History Theory and Instrumentation ...
Страница 34: ...34 RamanFlex 400 Series Getting Started Guide ...
Страница 35: ...Preparing your Facility ...
Страница 40: ...40 RamanFlex 400 Series Getting Started Guide ...
Страница 41: ...Installing your RamanFlex 400 ...
Страница 61: ...Introduction to the RamanFlex 400 ...
Страница 68: ...68 RamanFlex 400 Series Getting Started Guide ...
Страница 69: ...Using the RamanFlex 400 ...
Страница 85: ...Maintenance and Troubleshooting ...