Maintenance and Troubleshooting . 87
Cleaning the Air Filters
The air filters are located inside the top panel of the spectrometer module. You should clean
the filters regularly.
To clean the filters:
1.
Switch off the instrument power supply unit.
2.
Disconnect the line power cord.
3.
Remove the 4 screws holding the top cover in position and then lift off the cover.
Figure 19 Removing the top cover of the spectrometer module
4.
Unclip the filter covers, remove the filters and wash in a mild detergent.
5.
Dry the filters thoroughly.
6.
Fit the air filters back into the spectrometer module, and replace the covers.
7.
Place the top cover in position and replace the 4 screws.
Screws
Screws