Clarus 500 GC Installation Guide
99
7.
Connect the other end of the fused-silica tubing to the PID exit tubing as follows:
a)
Insert the nut on one end of the 1/16 to 1/16-in. union (P/N 0990-3075) over the
PID exit tube until it bottoms. Using two wrenches, tighten the nut 1/4 turn past
fingertight.
b)
Remove the nut and metal ferrule from the other end of the 1/16 to 1/16-in.
union. Put the metal ferrule aside; you will not need it.
c)
Insert the nut and a 1/16-inch graphite/Vespel ferrule (P/N 0992-0301) over the
free end of the fused-silica tubing.
d)
Insert the end of the fused-silica tubing into the union until it bottoms. Slide the
nut up to the union and tighten it fingertight.
e)
Using two 1/4-in. wrenches, tighten the nut only until the fused-silica tubing
cannot be pulled out of the ElCD fitting.
Do not overtighten the nut!
You can operate the ElCD and the PID in series or operate either detector independently.
Always connect your analytical column to the PID.
Figure 31. Connecting the fused silica line to the ElCD and the PID.
Содержание clarus 500 gc
Страница 1: ...Clarus 500 GC Installation Guide ...
Страница 5: ...Introduction 1 ...
Страница 6: ......
Страница 11: ...Clarus 500 GC Installation Guide 11 Warning risk of electric shock Warning refer to accompanying documents ...
Страница 12: ...Introduction 12 Label locations on instrument ...
Страница 13: ...Safety Practices 2 ...
Страница 14: ......
Страница 33: ...Preparing Your Laboratory 3 ...
Страница 34: ......
Страница 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Страница 45: ...Installing the Clarus 500 GC 4 ...
Страница 46: ......
Страница 58: ...Installing the Clarus 500 GC 58 ...
Страница 59: ...Connecting the Gases and Electrical Supply 5 ...
Страница 60: ......
Страница 67: ...Clarus 500 GC Installation Guide 67 Figure 12 Location of the Clarus 500 GC bulkheads on the rear panel ...
Страница 71: ...Clarus 500 GC Installation Guide 71 Figure 13 TCD in Channel B rear channel with one packed injector ...
Страница 79: ...Clarus 500 GC Installation Guide 79 Figure 21 ECD in Channel A PID ElCD in Channel B with two packed injectors ...
Страница 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Страница 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Страница 88: ...Connecting the Gases and Electrical Supply 88 ...
Страница 89: ...Connecting the Accessories 6 ...
Страница 90: ......
Страница 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Страница 105: ...Clarus 500 GC Installation Guide 105 Figure 34 Routing recorder integrator wires around the strain relief posts ...
Страница 106: ...Connecting the Accessories 106 ...
Страница 107: ...PPC Restrictor Information 7 ...
Страница 108: ......