Safety Practices
24
ECD Radioactive Hazards
WARNING
To assure that removable radioactive contamination on the external
parts of the ECD remains at a safe level, the United States Nuclear
Regulatory Commission requires that:
•
The ECD be wipe tested at least once every six months.
•
A record of the results must be maintained for NRC inspection.
United States Government Regulations for ECDs
NOTE: To repair an Electron Capture Detector cell requires a specific license issued by the U.S.
Nuclear Regulatory Commission (NRC) and/or in some states by the equivalent state agency.
For further information on obtaining a license, contact the Customer Service Department at
PerkinElmer, Shelton, Connecticut, or the NRC Material Branch, Office of Nuclear
Materials, Safety and Safeguards, Washington, DC 20555.
All USNRC regulations can be obtained through the internet at www.nrc.gov/reading-rm/
NOTE: These instructions are for ECD cell purchasers who are not specifically licensed to handle
radioactive materials.
The Clarus 500 GC Electron Capture Detector model (P/N N610-0063) contains a maximum
of 15 mCi of Nickel 63 (Ni 63), a radioactive material. Your possession and use of this
detector is governed by 10 C.F.R. Section 31.5 which is reproduced in Appendix I. Under the
provisions of that regulation you are deemed a "General Licensee."
Your possession and use of the detector cell may also be regulated by the state where you are
located. The requirements of state regulatory agencies are substantially similar to those
contained in NRC regulation 10 C.F.R. Section 31.5, but they may differ in some respects. It
is suggested that you procure a copy of the regulations of your particular state. (Supplement 2
in Appendix I contains a list of the "Agreement States" which have been granted authority by
the U.S. Nuclear Regulatory Commission to regulate the possession and use of radioactive
material.)
It is required that you be familiar with regulation 10 C.F.R. Section 31.5 (Appendix 1 in the
Hardware Guide 0993-6590). Following are summaries of its requirements.
Labels
Do not remove any of the labels attached to the ECD cell or any of the labels attached to your
Clarus 500 Gas Chromatograph that refer to the ECD cell. Follow all instructions and abide
by all precautions provided by the labels and in user instruction manuals referred to by the
labels.
Содержание clarus 500 gc
Страница 1: ...Clarus 500 GC Installation Guide ...
Страница 5: ...Introduction 1 ...
Страница 6: ......
Страница 11: ...Clarus 500 GC Installation Guide 11 Warning risk of electric shock Warning refer to accompanying documents ...
Страница 12: ...Introduction 12 Label locations on instrument ...
Страница 13: ...Safety Practices 2 ...
Страница 14: ......
Страница 33: ...Preparing Your Laboratory 3 ...
Страница 34: ......
Страница 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Страница 45: ...Installing the Clarus 500 GC 4 ...
Страница 46: ......
Страница 58: ...Installing the Clarus 500 GC 58 ...
Страница 59: ...Connecting the Gases and Electrical Supply 5 ...
Страница 60: ......
Страница 67: ...Clarus 500 GC Installation Guide 67 Figure 12 Location of the Clarus 500 GC bulkheads on the rear panel ...
Страница 71: ...Clarus 500 GC Installation Guide 71 Figure 13 TCD in Channel B rear channel with one packed injector ...
Страница 79: ...Clarus 500 GC Installation Guide 79 Figure 21 ECD in Channel A PID ElCD in Channel B with two packed injectors ...
Страница 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Страница 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Страница 88: ...Connecting the Gases and Electrical Supply 88 ...
Страница 89: ...Connecting the Accessories 6 ...
Страница 90: ......
Страница 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Страница 105: ...Clarus 500 GC Installation Guide 105 Figure 34 Routing recorder integrator wires around the strain relief posts ...
Страница 106: ...Connecting the Accessories 106 ...
Страница 107: ...PPC Restrictor Information 7 ...
Страница 108: ......