Clarus 500 GC Installation Guide
41
Power Specification:
All electrical supplies must be smooth, clean, and free of
line transients greater than 40 V peak to peak and must
meet and remain within the following tolerances:
120 VAC ±10% @ 50/60 Hz ±1%
230 VAC ±10% @ 50/60 Hz ±1%
Instruments and peripherals should not be connected to
circuits with large inductive or large and frequent loads
(i.e., large motors, discharge lamps, photocopy systems,
radio transmitters, etc.).
Power Outlets:
A minimum of one dedicated 120 VAC outlet at 20 A or
one 230 VAC outlet at 10 A (minimum) is required for
the GC. Additional equipment, such as computers and
printers, should be connected per their specifications.
Safety Requirements:
Gas Cylinders and
Gas Delivery Lines:
All gas cylinders should be firmly clamped to a
suitable surface. Care must be taken not to
kink or overstress the gas delivery lines.
Hydrogen:
Ensure that all hydrogen lines and connections
are leak-free. When using a hydrogen tank,
install an in-line hydrogen snubber (P/N 0009-
0038) between the tank regulator and the
delivery tubing.
Ventilation:
Always provide adequate ventilation. When
analyzing hazardous compounds, such as
pesticides, it may be necessary to arrange for
venting the detector effluent into a fume hood.
Содержание clarus 500 gc
Страница 1: ...Clarus 500 GC Installation Guide ...
Страница 5: ...Introduction 1 ...
Страница 6: ......
Страница 11: ...Clarus 500 GC Installation Guide 11 Warning risk of electric shock Warning refer to accompanying documents ...
Страница 12: ...Introduction 12 Label locations on instrument ...
Страница 13: ...Safety Practices 2 ...
Страница 14: ......
Страница 33: ...Preparing Your Laboratory 3 ...
Страница 34: ......
Страница 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Страница 45: ...Installing the Clarus 500 GC 4 ...
Страница 46: ......
Страница 58: ...Installing the Clarus 500 GC 58 ...
Страница 59: ...Connecting the Gases and Electrical Supply 5 ...
Страница 60: ......
Страница 67: ...Clarus 500 GC Installation Guide 67 Figure 12 Location of the Clarus 500 GC bulkheads on the rear panel ...
Страница 71: ...Clarus 500 GC Installation Guide 71 Figure 13 TCD in Channel B rear channel with one packed injector ...
Страница 79: ...Clarus 500 GC Installation Guide 79 Figure 21 ECD in Channel A PID ElCD in Channel B with two packed injectors ...
Страница 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Страница 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Страница 88: ...Connecting the Gases and Electrical Supply 88 ...
Страница 89: ...Connecting the Accessories 6 ...
Страница 90: ......
Страница 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Страница 105: ...Clarus 500 GC Installation Guide 105 Figure 34 Routing recorder integrator wires around the strain relief posts ...
Страница 106: ...Connecting the Accessories 106 ...
Страница 107: ...PPC Restrictor Information 7 ...
Страница 108: ......