Figure 30. Examples of a good tubing cut and bad cuts.
4.
Position the nut on the fused-silica tubing so that the back of the nut is
8.8 cm (3
1/2
inches) from the end of the fused-silica tubing. This positions the fused-
silica tubing inside the ElCD receiver.
5.
Using typewriter "white-out" or a felt-tipped pen, make a mark on the fused-silica
tubing just beyond the back edge of the nut.
CAUTION
To avoid contaminating the system, make sure the nut does not come
into contact with the mark on the fused-silica tubing.
6.
Connect the fused-silica tubing to the ElCD fitting inside the oven as follows:
a)
Insert this end of the fused-silica tubing into the ElCD fitting. Slide the nut up to
the ElCD and turn it until it is fingertight.
b)
Push the fused-silica tubing into the ElCD fitting until the mark on the tubing is
aligned with the back of the nut.
c)
Using two 1/4-inch wrenches, tighten the nut only until the fused-silica tubing
cannot be pulled out of the ElCD fitting.
Do not overtighten the nut!
Содержание clarus 500 gc
Страница 1: ...Clarus 500 GC Installation Guide ...
Страница 5: ...Introduction 1 ...
Страница 6: ......
Страница 11: ...Clarus 500 GC Installation Guide 11 Warning risk of electric shock Warning refer to accompanying documents ...
Страница 12: ...Introduction 12 Label locations on instrument ...
Страница 13: ...Safety Practices 2 ...
Страница 14: ......
Страница 33: ...Preparing Your Laboratory 3 ...
Страница 34: ......
Страница 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Страница 45: ...Installing the Clarus 500 GC 4 ...
Страница 46: ......
Страница 58: ...Installing the Clarus 500 GC 58 ...
Страница 59: ...Connecting the Gases and Electrical Supply 5 ...
Страница 60: ......
Страница 67: ...Clarus 500 GC Installation Guide 67 Figure 12 Location of the Clarus 500 GC bulkheads on the rear panel ...
Страница 71: ...Clarus 500 GC Installation Guide 71 Figure 13 TCD in Channel B rear channel with one packed injector ...
Страница 79: ...Clarus 500 GC Installation Guide 79 Figure 21 ECD in Channel A PID ElCD in Channel B with two packed injectors ...
Страница 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Страница 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Страница 88: ...Connecting the Gases and Electrical Supply 88 ...
Страница 89: ...Connecting the Accessories 6 ...
Страница 90: ......
Страница 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Страница 105: ...Clarus 500 GC Installation Guide 105 Figure 34 Routing recorder integrator wires around the strain relief posts ...
Страница 106: ...Connecting the Accessories 106 ...
Страница 107: ...PPC Restrictor Information 7 ...
Страница 108: ......