Clarus 500 GC Installation Guide
85
In Countries with 230V Single-Phase Power
WARNING
Only a PerkinElmer Service representative or similarly trained and
authorized personnel should ever change or replace the plug for the
Clarus 500 GC.
1.
Select the appropriate plug and connect it to the three power cord wires as follows:
•
Connect the brown wire to line (AC).
•
Connect the blue wire to neutral (ACC).
•
Connect the green/yellow wire to chassis ground.
2.
Remove the three screws that secure the right side panel to the Clarus 500 GC. Then
remove the panel.
3.
Verify that the brown wire from the line filter is connected to E1, that the blue wire
from the line filter is connected to E2, and that the yellow/green wire is connected to
E3 on the AC Distribution P.C. board (see Figure 23).
4.
Make sure that a 10-A Slow-Blow fuse is installed in the F3 fuse holder and a wire is
soldered across the contacts of the F3 fuse holder to short this connection.
5.
Replace the right side panel and connect the power cord to a suitable electrical outlet.
In Countries with 230V Two-Phase Power
WARNING
Only a PerkinElmer Service representative or similarly trained and
authorized personnel should ever change or replace the plug for the
Clarus 500 GC.
1.
Select the appropriate plug and connect it to the three power cord wires as follows:
Содержание clarus 500 gc
Страница 1: ...Clarus 500 GC Installation Guide ...
Страница 5: ...Introduction 1 ...
Страница 6: ......
Страница 11: ...Clarus 500 GC Installation Guide 11 Warning risk of electric shock Warning refer to accompanying documents ...
Страница 12: ...Introduction 12 Label locations on instrument ...
Страница 13: ...Safety Practices 2 ...
Страница 14: ......
Страница 33: ...Preparing Your Laboratory 3 ...
Страница 34: ......
Страница 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Страница 45: ...Installing the Clarus 500 GC 4 ...
Страница 46: ......
Страница 58: ...Installing the Clarus 500 GC 58 ...
Страница 59: ...Connecting the Gases and Electrical Supply 5 ...
Страница 60: ......
Страница 67: ...Clarus 500 GC Installation Guide 67 Figure 12 Location of the Clarus 500 GC bulkheads on the rear panel ...
Страница 71: ...Clarus 500 GC Installation Guide 71 Figure 13 TCD in Channel B rear channel with one packed injector ...
Страница 79: ...Clarus 500 GC Installation Guide 79 Figure 21 ECD in Channel A PID ElCD in Channel B with two packed injectors ...
Страница 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Страница 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Страница 88: ...Connecting the Gases and Electrical Supply 88 ...
Страница 89: ...Connecting the Accessories 6 ...
Страница 90: ......
Страница 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Страница 105: ...Clarus 500 GC Installation Guide 105 Figure 34 Routing recorder integrator wires around the strain relief posts ...
Страница 106: ...Connecting the Accessories 106 ...
Страница 107: ...PPC Restrictor Information 7 ...
Страница 108: ......