Clarus 500 GC Installation Guide
69
Common Injector/Detector Gas Connections
Figures 13 through 21 shows how gases are connected to the Clarus 500 GC in several
common injector/detector configurations. Note that the strain relief, which is required on all
gas lines, is not shown in these figures.
WARNING
Position incoming gas lines away from the oven exhaust duct so that the hot
exhaust will not blow directly on the gas lines.
CAUTION
Thoroughly purge the incoming gas lines before measuring and adjusting the
gases. This is especially important when using hydrogen gas.
NOTE
:
For Clarus 500 GC’s that have a combination of both PPC and manual pneumatics, set the
detector gas delivery pressure between 60 and 90 psig and install the fixed 30-psig stainless
steel diaphragm regulator (P/N N610-1473), behind the Clarus 500 GC for the manual
pneumatics detector gases.
The manual pneumatics bulkhead always contains Swagelok nuts in all six locations whether
or not there is an injector or a detector connected to this bulkhead. The PPC modules will be
installed only if your Clarus 500 GC has been configured with PPC for an injector and/or
detector.
The Clarus 500 GC can be configured as any of the following:
•
All zones (injector/detector) as PPC.
•
All zones as manual pneumatics.
Содержание clarus 500 gc
Страница 1: ...Clarus 500 GC Installation Guide ...
Страница 5: ...Introduction 1 ...
Страница 6: ......
Страница 11: ...Clarus 500 GC Installation Guide 11 Warning risk of electric shock Warning refer to accompanying documents ...
Страница 12: ...Introduction 12 Label locations on instrument ...
Страница 13: ...Safety Practices 2 ...
Страница 14: ......
Страница 33: ...Preparing Your Laboratory 3 ...
Страница 34: ......
Страница 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Страница 45: ...Installing the Clarus 500 GC 4 ...
Страница 46: ......
Страница 58: ...Installing the Clarus 500 GC 58 ...
Страница 59: ...Connecting the Gases and Electrical Supply 5 ...
Страница 60: ......
Страница 67: ...Clarus 500 GC Installation Guide 67 Figure 12 Location of the Clarus 500 GC bulkheads on the rear panel ...
Страница 71: ...Clarus 500 GC Installation Guide 71 Figure 13 TCD in Channel B rear channel with one packed injector ...
Страница 79: ...Clarus 500 GC Installation Guide 79 Figure 21 ECD in Channel A PID ElCD in Channel B with two packed injectors ...
Страница 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Страница 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Страница 88: ...Connecting the Gases and Electrical Supply 88 ...
Страница 89: ...Connecting the Accessories 6 ...
Страница 90: ......
Страница 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Страница 105: ...Clarus 500 GC Installation Guide 105 Figure 34 Routing recorder integrator wires around the strain relief posts ...
Страница 106: ...Connecting the Accessories 106 ...
Страница 107: ...PPC Restrictor Information 7 ...
Страница 108: ......