Preparing Your Laboratory
40
Environmental Requirements:
Pollution Degree:
This instrument will operate safely in environments
that contain nonconductive foreign matter up to
Pollution Degree 2 as defined in IEC 1010-1.
Laboratory Environment:
Install the GC in an indoor laboratory environment
that is clean and is free of drafts, direct sunlight and
vibration.
The laboratory should be free of flammable,
explosive, toxic, caustic, or corrosive vapors or
gases and should be relatively free of dust.
The ambient laboratory temperature should be
between 10 °C and 35 °C (50 °F and 95 °F) with a
relative humidity between 20% and 80% with no
condensation. The GC will operate safely between 5
°C and 40 °C (41 °F and 104 °F).
Electrical Power Requirements:
Installation Category Statement:
This instrument is able to withstand transient
over-voltage according to Installation
Category II as defined in IEC 1010-1.
Power Consumption:
2 400 VA (volt-amps) for the GC. Add
100 VA for the computer and 108 VA for the
printer.
Содержание clarus 500 gc
Страница 1: ...Clarus 500 GC Installation Guide ...
Страница 5: ...Introduction 1 ...
Страница 6: ......
Страница 11: ...Clarus 500 GC Installation Guide 11 Warning risk of electric shock Warning refer to accompanying documents ...
Страница 12: ...Introduction 12 Label locations on instrument ...
Страница 13: ...Safety Practices 2 ...
Страница 14: ......
Страница 33: ...Preparing Your Laboratory 3 ...
Страница 34: ......
Страница 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Страница 45: ...Installing the Clarus 500 GC 4 ...
Страница 46: ......
Страница 58: ...Installing the Clarus 500 GC 58 ...
Страница 59: ...Connecting the Gases and Electrical Supply 5 ...
Страница 60: ......
Страница 67: ...Clarus 500 GC Installation Guide 67 Figure 12 Location of the Clarus 500 GC bulkheads on the rear panel ...
Страница 71: ...Clarus 500 GC Installation Guide 71 Figure 13 TCD in Channel B rear channel with one packed injector ...
Страница 79: ...Clarus 500 GC Installation Guide 79 Figure 21 ECD in Channel A PID ElCD in Channel B with two packed injectors ...
Страница 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Страница 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Страница 88: ...Connecting the Gases and Electrical Supply 88 ...
Страница 89: ...Connecting the Accessories 6 ...
Страница 90: ......
Страница 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Страница 105: ...Clarus 500 GC Installation Guide 105 Figure 34 Routing recorder integrator wires around the strain relief posts ...
Страница 106: ...Connecting the Accessories 106 ...
Страница 107: ...PPC Restrictor Information 7 ...
Страница 108: ......