![PCE Instruments PCE-RTD 20 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/pce-instruments/pce-rtd-20/pce-rtd-20_user-manual_1516150035.webp)
© PCE Instruments
33
4.6.2
Display Settings
Display
Intensity
Display Brightness Settings.
Range:
5 to 100
Display Off
Time
Standby Time in second after which
display will turn Off. To turn the
display off press any key.
Range:
0 to 9999 sec
Setting 0 will disable this feature.
That means display will never turn off
automatically.
4.6.3
Date/Time Settings
To set the Time + Date of the device.
2 date formats are supported in this device:
DD/MM/YY
and
MM/DD/YY
. This is useful only in Data Logging, to
decide in which format the Date is to be stored.
2 time formats are supported in this device:
24 Hour
and
12 Hour
. This is to select in which format the time should
be displayed on Run page and time to be stored in Data
Logging.
AM/PM selection is enabled only for 12 Hour Time Format.
4.6.4
Calibration
The instrument is factory calibrated for the specified range, but due to long term drift of
components, re-calibration may be necessary in some cases. For calibrating the instrument, a
reliable source is required. This source should be at least ten times accurate compared to the
range of the instrument.
Note
: PCE Instruments can provide a calibration service that is traceable to international
standards.
We recommend that you return the instrument to the manufacturer or an approved service agent
for calibration. If you use an alternative calibration facility, make sure that it uses these standards.