Parameter
Name
Description / Options
Main Display
Select which Reading to be display as a Main Reading (Reading
Displays in Box in RUN Page).
Available Options:
Actual
Display the Actual Input Value
Percentage
Display the Percentage Value of the Input
The Value depends on Input Range. These
settings are available from
MENU
→
EM SETUP.
Scaled
Display the Scaled Value of the Input
The Scale Value depends on Input Range, Input
Scaled Range and Transfer Function. These
settings are available from
MENU
→
EM SETUP.
Filter(sec)
1
st
Order IIR Low Pass Filter for Input Reading
Filter is useful when a measurement signal contains unwanted noise.
Range:
0.0 … 60.0 sec
Tare(unit)
The Tare value is subtracted from the reading of the measured value.
Here unit is changed according to current Input Type and Measure
Display Mode.
Range:
In accordance with Input Range and Measure Display Mode.
Note: Beware of the problems that may result in not seeing the
true measurement value
.
4.4.2
RTD Display Settings
This Page is appears in RUN
→
MENU
→
DISPLAY
→
RTD Terminal.