![PCE Instruments PCE-RTD 20 Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/pce-instruments/pce-rtd-20/pce-rtd-20_user-manual_1516150027.webp)
© PCE Instruments
25
4.3.3
Continuity Test
Parameter Name
Description / Options
Continuity
Feature
Continuity Feature Selection RadioButtonBox
Selecting RadioButtonBox
will enable continuity feature for RTD
Measure Mode.
Threshold
Enter the threshold value of resistance up to which continuity test is
applied.
Range:
0 … 100
When Testing Continuity Beep sounds and continuity symbol appear on run page as shown in
below figure. When resistance between the
Ω
Measure terminal is less than 25
Ω
(or defined in
threshold parameter). To test the continuity remove power from the circuit to be tested.