![PCE Instruments PCE-RTD 20 Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/pce-instruments/pce-rtd-20/pce-rtd-20_user-manual_1516150012.webp)
© PCE Instruments
10
3.2
User Interface
Everytime the PCE-RTD 20 is switched on, the startup message ends in RUN Page.
3 display modes are available in RUN Page.
1. RTD Measure/Source Mode
2. EM Measure Mode
3. RTD Measure/ EM Measure Mode
This Display Mode can be selected from MENU
→
DISPLAY Page.
In case of Dual Mode Display Screen is divided into two parts. Due to that only few additional
info will appear on RUN Page. But Which Information to be shows can be selectable in Display
Mode Menu.
3.2.1
Status Bar
The status bar at the top of the display is visible only in RUN Page. It is divided into five main
sections.
1
Time in HH:MM:SS format
Available in two formats
1. 24 hour (default)
2. 12 hour
This setting is available in Date/Time in Settings Menu
2
Error Code Indicator
This icon is visible if any on-board peripherals like RTC, ADC, DAC, etc. are not
working properly, see chapter 5 Maintenance & Troubleshooting.
The List of error codes available in this device is given below.
Error code
Description
0
Memory corrupted or device unable to read/write it
1
RTC not working properly
2
Device unable to read battery information
3
Measure mode not working
4
Device unable to get source feedback reading
5
Data log memory corrupt
6
Source mode not working
9
More than one error from above list is occurring