Manual Stepping
To Enable Manual Stepping, select Source Type as STEP and Check the Manual CheckBox.
If this mode is enabled,
icon will appear in Source Display Window in RUN Page.
Pressing UP or DOWN key in RUN Page will Increment or Decrement Source Value by Step
specified in STEP Page.
In RUN Page, Source Value can directly change by Pressing
F1
key (EDIT) and modifying value
like in EditBox. And STEP Setting can be accessed directly by
F2
key (SETTING).
Auto Stepping
To Enable Auto Stepping, select Source Type as STEP and Un-Check the Manual CheckBox.
If this mode is enabled,
(Step UP) or
(Step Down) icon will appear in Source Display
Window in RUN Page and F1 and F2 Button change to
START
and
SETTING
respectively.
Automated Step can be started by Pressing
F1
key (START). After that F1 and F2 key will
change to
PAUSE
and
STOP
respectively. So by pressing F1 and F2 key running STEP can be
PAUSE or STOP at any time in RUN Page.
STEP Setting can be accessed directly by
F2
key (SETTING).
Note
: While STEP is running, STEP settings are not accessible and Source Page Parameter
settings can’t be changed. Stop STEP before changing any settings.
4.3.2
RAMP Page
Parameter Name
Description / Options
Low
Starting Value of Ramp
Enter value according to RTD Display Mode. If display mode is actual
enter value in ohms and if display mode is %, enter value in %.
High
Ending Value of Ramp
Enter value according to RTD Display Mode. If display mode is actual
enter value in ohms and if display mode is %, enter value in %.
Hold@0%(s)
Time to wait at Low (0%) level in second
This parameter is used for
Repeat Format UP/DOWN
or
DOWN/UP
.
Range:
0 … 9999