© PCE Instruments
29
Available Options for RTD Source Mode:
Options
Icon
Description
None
-
No info is visible.
Actual Value
Shows the Actual RTD
Temperature/ohms value without
any scaling.
This option is available only if RTD
Display Mode is Percentage.
Excitation
Current
Shows the current which is
sourced by instrument under test.
Filter(sec)
1
st
Order IIR Low Pass Filter for RTD Measure Reading.
This option is available only for RTD mode is Measure.
Filter is useful when a measurement signal contains unwanted noise.
Range:
0.0 … 60.0 sec
4.5
DATA LOGGING Page
This section gives examples of how to log Readings with time and date over a set time period or
on a key press. Logged data is stored in a user defined file in internal memory.
This Page is appears in
RUN
→
MENU
→
LOGGING.
Parameter Name
Description / Options
Trigger
Data Logging Trigger Mode Selection
Available Options:
Key Press
Log Data on pressing
from RUN key Page.
Periodic
Log Data periodically at every Sampling Rate for
total time specified by Logging Time.