background image

Manual 

 

Example: 

 

-> upper limit -> set point = -200 N 

                                                   

Sign = absolute value 

 
In both cases, when + 200 N as well as -200 N is exceeded the upper limit is active. 
 
                 

 

-> upper limit -> set point = -200 N 

                                                     

Sign = to be signed 

Only when -200 N is exceeded, the upper limit is active. +

 

5.2.2.2 

Time limit for data transfer 

 

In the menu item 

"DSI time limit activation"

 you can activate a time limit for stopping the data 

transmission. 
 
The menu item 

"DSI time limit (sec)"

 allows you to enter certain time that represents how long the data 

will transferred via USB interface. The input is made in seconds (0 ... 9999 s). If the time is set to 0, this 
function is disabled. 
 

Note:

 The limit value outputs of the device are switched when the overruns of the time limit happen. A 

connected test stand is thereby stopped. To reset the outputs, press either 

 or start a new data 

transfer. 
 

If the function "Save the curve" is activated, this setting is not effective. Instead, the time limit is 
considered from "curve duration". 

5.2.3 

Display 

 
You can adjust display as follows according to your needs: 

 

"Rotate display " 

After pressing

 

 you can rotate the display by 180 ° with the help of

. Confirm with 

.. 

 

"Background"

 

After pressing 

 you can select between 

Normal

 and 

Inverted

 display with the help of 

..Confirm 

with 

 

 

"Backlight"

 

After pressing 

 the backlight can be controlled with 

 smoothly. Confirm the selection with 

.. 

 

"Contrast"

 

After pressing 

 the contrast can be controlled with 

.smoothly. Confirm the selection with 

 

 

5.2.4 

TEDs (automatic sensor recognition) 

Information:

 what is TEDs? TEDs (transducer electronic data sheet) 

is an international standard for 

sensors. Each TEDs measuring cell has a memory chip integrated in the sensor or plug, on which the 
basic information on the measuring cell is stored. These are, for example, serial number, measurement 
range or calibration. 
When a TEDs measuring cell is connected to the meter, you can view the information stored on the 
memory chip in this menu.

 

 
"Reading  ID" 

 

By pressing 

 a new window is opened, which contains the basic TEDs information. Navigate here with 

Содержание PCE-MMT E

Страница 1: ...d Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce instruments co...

Страница 2: ...ilter 13 5 4 Factory setup 13 6 Save and transfer the measurement data 13 6 1 Peak values 13 6 2 Curve 14 6 3 Termination of dara saving 14 6 4 Data transfer in mass storage mode 14 6 5 Data transfer...

Страница 3: ...are at any time and without notice Any changes will be reflected in the then published user manuals 2 Safety notes Please read this manual carefully and completely before you use the device for the fi...

Страница 4: ...1 Key functions 4 5 6 7 1 2 3 1 Power button To turn the device on please press the button for 2 3 seconds Then the device starts and displays the connected TEDs measuring cells Notes on TEDs in chap...

Страница 5: ...er mode or the curve recording in a mass storage mode 5 Mode Measure without peak value Zeroing of the indicated force Mode Menu One menu level higher from the submenu Input of the numerical value mov...

Страница 6: ...he mode you need for measurement In the menu item Settings you can carry out the expanded settings The menu item Factory Setup makes it possible to restore the factory default settings or to enter the...

Страница 7: ...values are retained and displayed again when the peak measurement is enabled They will be only overwritten with higher values Curve storage Curve Clicking activates the function curve storage 5 2 1 1...

Страница 8: ...er distance limit value can be set in addition Note The data transfer and storage is automatically stopped when a limit value is exceeded Two symbols on the display above the power indicator show the...

Страница 9: ...d the time limit is considered from curve duration 5 2 3 Display You can adjust display as follows according to your needs Rotate display After pressing you can rotate the display by 180 with the help...

Страница 10: ...ched on measuring instrument as well as when the device is switched on the detected measurement cell information is displayed A new window open which displays the main information of the measurement c...

Страница 11: ...y only in case of battery pack change as the date and time are set at the factory With you can set the time and date The input is activated after the initial pressing of any of both keys Confirm with...

Страница 12: ...the function for the automatic zeroing when switching By pressing you confirm the selection When this function is active the display is set to zero at switching automatically regardless of the existi...

Страница 13: ...filter are though the computational complexity is also higher then For the display device it means that the higher the order of the selected filter the more values are taken into account in the avera...

Страница 14: ...menu item no saving see chapter or when the memory on the SD card is completely full The measurement data transfer through the data streaming remains active 6 4 Data transfer in mass storage mode Data...

Страница 15: ...er switching between force time and force distance curves Note Please always pay attention that the device must be turned on first before it is connected to the started computer If the data transfer d...

Страница 16: ...batteries is shown at any time in the upper right corner of the display If the batteries have reached their maximum service life and no longer have their full power so standard AA batteries or in case...

Страница 17: ...bout 15 a header line Overload appears instead of the force display A mechanical overload protection happens at about 125 of rated load If the overload is withdrawn in time this indication disappears...

Страница 18: ...Lower 1221 Setpoint Input 1222 Mode Sign Absolute 13 Display 131 Rotate 1311 0 180 132 Invert 1321 Normal Inverted 133 Backlight 1331 Power Off Max 134 Contrast 1341 Power Off Max 14 Units 141 N 142 k...

Страница 19: ...em or give them to a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 16 Contact If you have any questions about our range of p...

Отзывы: