Manual 

16 

Note

: How the data is transferred depends on the measuring frequency set on the device. With the set 

measuring frequency of 100 Hz, the data is transferred directly and continuously from the start of the 
measurement. If the measured frequency is above 100 Hz, the measurement data is first cached in the 
meter due to a limitation of the USB interface and then transferred at the end of the measurement. 

 

Calibration 

 
 

The calibration of the display devices and sensors is done in a certified laboratory using traceable 
calibrated weights (up to 1000 N) or traceable calibrated reference-force gauges (1000 N). 
It is recommended to carry out and certify recalibration annually. 
 

Battery life and charging 

 
 

In the compartment on the back side of the meter there are three environment friendly AA battery packs 
that are fully charged before delivery. To charge the batteries, the USB power adapter supplied should be 
used. Please work several times until the complete discharge of the batteries and charge them again only 
after that. In case of lasting use, for example, in small tests, disconnect the charger at least once a week 
during working hours and wait until the full discharge of the batteries. Please charge the device only after 
that again. The normal working hours with a fully charged battery pack is about 10 hours. The necessary 
charging time is approximately 3 - 4 hours. The current state of charge of the batteries is shown at any 
time in the upper right corner of the display. If the batteries have reached their maximum service life and 
no longer have their full power, so standard AA batteries or in case of emergency 1.5V AA batteries can 
be used. Please note that the working time with standard batteries is limited. When connecting the 
delivered charger you can see the state of charge by the moving battery icon in the upper right corner of 
the display. Please note that the new batteries reach their full capacity only after a few charge / discharge 
cycles. It is recommended to connect the handheld instrument to the charger after each use or daily. 
Note: It makes sense to discharge the batteries after long use as much as possible and to charge them 
fully again. 

 

SD memory card 

 

A memory card with a capacity of 2 GB is included in the delivery. On it, the meter stores accumulating 
peak values and curve files. Alternatively, for data transmission via the USB cable, this card with the help 
of a small pressure can be taken out of the slot located on the right side. With the help of a SD card 
reader it can be read on any computer. SD memory cards up to 4GB are supported.

 

 

10  In- and outputs 

10.1  Mini-USB interface 

 
The USB-data cable is included in the delivery. 
 

10.2  Output 

 

The analog output is proportional to the measurement value. This has a max. value of ± 2 V DC, 
according to the nominal range of the hand-held meter. 
 

A possible calibration of the analog output at two measuring points is available optionally. 

 
Note

: The zero adjustment on the display device is carried out in the unloaded condition of the load cell. 

Содержание PCE-MMT E

Страница 1: ...d Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce instruments co...

Страница 2: ...ilter 13 5 4 Factory setup 13 6 Save and transfer the measurement data 13 6 1 Peak values 13 6 2 Curve 14 6 3 Termination of dara saving 14 6 4 Data transfer in mass storage mode 14 6 5 Data transfer...

Страница 3: ...are at any time and without notice Any changes will be reflected in the then published user manuals 2 Safety notes Please read this manual carefully and completely before you use the device for the fi...

Страница 4: ...1 Key functions 4 5 6 7 1 2 3 1 Power button To turn the device on please press the button for 2 3 seconds Then the device starts and displays the connected TEDs measuring cells Notes on TEDs in chap...

Страница 5: ...er mode or the curve recording in a mass storage mode 5 Mode Measure without peak value Zeroing of the indicated force Mode Menu One menu level higher from the submenu Input of the numerical value mov...

Страница 6: ...he mode you need for measurement In the menu item Settings you can carry out the expanded settings The menu item Factory Setup makes it possible to restore the factory default settings or to enter the...

Страница 7: ...values are retained and displayed again when the peak measurement is enabled They will be only overwritten with higher values Curve storage Curve Clicking activates the function curve storage 5 2 1 1...

Страница 8: ...er distance limit value can be set in addition Note The data transfer and storage is automatically stopped when a limit value is exceeded Two symbols on the display above the power indicator show the...

Страница 9: ...d the time limit is considered from curve duration 5 2 3 Display You can adjust display as follows according to your needs Rotate display After pressing you can rotate the display by 180 with the help...

Страница 10: ...ched on measuring instrument as well as when the device is switched on the detected measurement cell information is displayed A new window open which displays the main information of the measurement c...

Страница 11: ...y only in case of battery pack change as the date and time are set at the factory With you can set the time and date The input is activated after the initial pressing of any of both keys Confirm with...

Страница 12: ...the function for the automatic zeroing when switching By pressing you confirm the selection When this function is active the display is set to zero at switching automatically regardless of the existi...

Страница 13: ...filter are though the computational complexity is also higher then For the display device it means that the higher the order of the selected filter the more values are taken into account in the avera...

Страница 14: ...menu item no saving see chapter or when the memory on the SD card is completely full The measurement data transfer through the data streaming remains active 6 4 Data transfer in mass storage mode Data...

Страница 15: ...er switching between force time and force distance curves Note Please always pay attention that the device must be turned on first before it is connected to the started computer If the data transfer d...

Страница 16: ...batteries is shown at any time in the upper right corner of the display If the batteries have reached their maximum service life and no longer have their full power so standard AA batteries or in case...

Страница 17: ...bout 15 a header line Overload appears instead of the force display A mechanical overload protection happens at about 125 of rated load If the overload is withdrawn in time this indication disappears...

Страница 18: ...Lower 1221 Setpoint Input 1222 Mode Sign Absolute 13 Display 131 Rotate 1311 0 180 132 Invert 1321 Normal Inverted 133 Backlight 1331 Power Off Max 134 Contrast 1341 Power Off Max 14 Units 141 N 142 k...

Страница 19: ...em or give them to a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 16 Contact If you have any questions about our range of p...

Отзывы: