Manual 

10 

the help of 

 
"Reading Sensor" 

 

By pressing 

 a new window is opened, which contains the physical information of the cell (for 

example, the measuring range). Navigate here with the help of 

.

 

"Reading Calibration" 

 

By pressing 

 a new window is opened, which contains the calibration information of the measuring 

cell. Navigate here with the help of 

.  

5.2.4.1 

Connection of a TEDs measuring cell 

 

Once a cell with TEDs memory chip is connected to a switched on measuring instrument, as well as when 
the device is switched on, the detected measurement cell information is displayed.  
 
A new window open which displays the main information of the measurement cell (manufacturer, model 
and measuring range. 

 

 

 
 

By pressing 

, you acknowledge that the calibration stored on the TEDs memory chip is loaded into the 

meter. Pressing 

 results in stopping the downloading and previously set user settings such as limit 

values and units remain retained. 
In the next window you can select one of eight memory locations on the meter. Here, the calibration is 

loaded to the connected measuring cell. Confirm your choice with 

 
 
 
 
 
 

 

Содержание PCE-MMT E

Страница 1: ...d Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce instruments co...

Страница 2: ...ilter 13 5 4 Factory setup 13 6 Save and transfer the measurement data 13 6 1 Peak values 13 6 2 Curve 14 6 3 Termination of dara saving 14 6 4 Data transfer in mass storage mode 14 6 5 Data transfer...

Страница 3: ...are at any time and without notice Any changes will be reflected in the then published user manuals 2 Safety notes Please read this manual carefully and completely before you use the device for the fi...

Страница 4: ...1 Key functions 4 5 6 7 1 2 3 1 Power button To turn the device on please press the button for 2 3 seconds Then the device starts and displays the connected TEDs measuring cells Notes on TEDs in chap...

Страница 5: ...er mode or the curve recording in a mass storage mode 5 Mode Measure without peak value Zeroing of the indicated force Mode Menu One menu level higher from the submenu Input of the numerical value mov...

Страница 6: ...he mode you need for measurement In the menu item Settings you can carry out the expanded settings The menu item Factory Setup makes it possible to restore the factory default settings or to enter the...

Страница 7: ...values are retained and displayed again when the peak measurement is enabled They will be only overwritten with higher values Curve storage Curve Clicking activates the function curve storage 5 2 1 1...

Страница 8: ...er distance limit value can be set in addition Note The data transfer and storage is automatically stopped when a limit value is exceeded Two symbols on the display above the power indicator show the...

Страница 9: ...d the time limit is considered from curve duration 5 2 3 Display You can adjust display as follows according to your needs Rotate display After pressing you can rotate the display by 180 with the help...

Страница 10: ...ched on measuring instrument as well as when the device is switched on the detected measurement cell information is displayed A new window open which displays the main information of the measurement c...

Страница 11: ...y only in case of battery pack change as the date and time are set at the factory With you can set the time and date The input is activated after the initial pressing of any of both keys Confirm with...

Страница 12: ...the function for the automatic zeroing when switching By pressing you confirm the selection When this function is active the display is set to zero at switching automatically regardless of the existi...

Страница 13: ...filter are though the computational complexity is also higher then For the display device it means that the higher the order of the selected filter the more values are taken into account in the avera...

Страница 14: ...menu item no saving see chapter or when the memory on the SD card is completely full The measurement data transfer through the data streaming remains active 6 4 Data transfer in mass storage mode Data...

Страница 15: ...er switching between force time and force distance curves Note Please always pay attention that the device must be turned on first before it is connected to the started computer If the data transfer d...

Страница 16: ...batteries is shown at any time in the upper right corner of the display If the batteries have reached their maximum service life and no longer have their full power so standard AA batteries or in case...

Страница 17: ...bout 15 a header line Overload appears instead of the force display A mechanical overload protection happens at about 125 of rated load If the overload is withdrawn in time this indication disappears...

Страница 18: ...Lower 1221 Setpoint Input 1222 Mode Sign Absolute 13 Display 131 Rotate 1311 0 180 132 Invert 1321 Normal Inverted 133 Backlight 1331 Power Off Max 134 Contrast 1341 Power Off Max 14 Units 141 N 142 k...

Страница 19: ...em or give them to a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 16 Contact If you have any questions about our range of p...

Отзывы: