P
AGE
45
Er8
- Cal2 Volt Input
1. Check that 100mVDC is properly connected to the
Error
instrument and is within the tolerance limts as
indicated in the CAL2 procedure of the Calibration
section (page 35).
2. Turn off the power to the instrument and open the
cover. Loosen the platen hold down screw and swing
the platen open. Inspect the Processor board
(Appendix A-1, page 46) to insure that the input
conditioning jumper JU4 is in the non-volt position.
The jumper must be in the non-volt position to perform
the
CAL2
procedure properly. Close the platen and
tighten the hold down screw. Close the cover and
turn on the power to the instrument.
3. Perform the
CAL2
procedure as described in the
Calibration section (page 35).
Er9
- ADC Reference
1. Perform the
CAL2
procedure as described in the
Number Error
Calibration section (page 35).
Er10
- ADC Reference
1. Perform the
CAL2
procedure as described in the
Voltage Error
Calibration section (page 35).
Er12
- CAL2 Voltage Error
1. Check that 100 mVDC is properly connected to the
instrument and is within the tolerance limits as
indicated in the
CAL2
procedure of the Calibration
section (page 35)
2. Turn off the power to the instrument and open the
cover. Loosen the platen hold down screw and swing
the platen open . Inspect the Processor board
(Appendix A-1, page 46) to insure that the input
conditioning jumper JU4 is in the non-volt position.
3. Perform the
CAL2
procedure as described in the
Calibration section (page 35).