Appendix D
Specifications
Measurement Error Limit
mA, mV, and VDC +/- 0.25% of scaled span plus 1 Least
Significant Digit @ 25 degrees C
Ambient Temperature Error
0.01% of span per degree C deviation from 25 degrees C
Scan Rate
1 to 5 scans/second (One pen units)
1 to 3 scans/second (Two pen units)
Display Decimal Positions
None, One, two or three decimal places
Noise Rejection
Normal mode, 85dB minimum at 60Hz or greater
Common mode, 90dB minimum, 115VAC maximum
Line Voltage
115//- 10% 50/60 Hz
Power Consumption
25VA maximum
Operating Temperature
0 to 55
°
C (ambient)
32 to 131
°
F
Storage Temperature
-40 to 65
°
C
-40 to 149
°
F
Humidity
0 to 90% RH, noncondensing
Dimensions
13.19" H x 15.13" W x 3.63" Deep
Weight
20 pounds maximum
Transmitter Fault Detection
Displays
SnSr
for sensor or transmitter break. Alarm
Outputs go off. Presets remain as they were, Pulsed
outputs stop.Fault detection is not functional for 0-5V
or 0-20mA inputs.
Agency Approvals
UL and CSA
Transmitter Power Supply
Provides up to 42mA of current at 24VDC
Warranty
3 years, see back page for details.