The alarm hysteresis value can be adjusted between 0 – 10% of the selected display range using the
keys.
Pressing
OK
key will enter the new value, pressing the
MENU
key cancels any changes made and exits from the
hysteresis set up.
10.4 Fault Alarm Setup
When the alarm type is set to fault, the alarm fault setup determines which faults the alarm output will signal (at least
one fault must be selected for the alarm output to work).
The faults that can be signalled by the alarm are displayed as a list, the enabled faults are indicated by small ticks. To
enable a fault, use the
keys to highlight the required fault and press the
OK
key. A small tick will appear next to the
selected fault. To disable a fault, use the
keys to select it and press the
OK
key, the small tick will disappear.
Note: The instrument faults are defined elsewhere within this manual.
10.5 Delay to Set
A delay can be introduced between conditions for an alarm turning on being met and the alarm actually turning on. At
any time during a delay, if the conditions for turning the alarm on disappear, then the delay will end and the alarm will
remain off (no change). The delay to set can be varied between 0-600 seconds using the
keys. Pressing
OK
will store the new value.
10.6 Delay to Reset
A delay can be introduced between conditions for an alarm turning off being met and the alarm actually turning off.
Such a delay can be varied between 0-600 seconds. At any time during a delay, if the conditions for turning the alarm
off disappear, then the delay will end and the alarm will remain on (no change).
10.7 Test Alarm Output
This facility enables the alarm output to be tested to enable verification of its operation and also to check that
equipment connected to it is working properly. By using the
keys the alarm output can be toggled between its
NORMAL (OFF) and ALARM (ON) conditions. Press the
OK
or
MENU
key to exit.
10.8 Alarm Test Cycle
The alarm test cycle can be used when installing and configuring remote equipment on the alarm output. The alarm
output is automatically toggled between its NORMAL (OFF) and ALARM (ON) conditions every fifteen seconds and
will continue to do so for fifteen minutes after which the instrument will return to the measure mode. To exit the alarm
test cycle, press the
OK
or
MENU
key.
167300IM-Issue 12 Issue Date 14/10/2014
Page 23 of 50
Off
On
Alarm
Hysteresis
Alarm Value
Measured Value
Low Alarm with hysteresis
Off
On
Alarm
Hysteresis
Alarm Value
Measured Value
High Alarm with hysteresis
Содержание 7200
Страница 1: ...INSTRUCTION MANUAL 7200 Monitor Field Mount...
Страница 2: ...This page is intentionally left blank...
Страница 26: ...7200 Monitor Instruction Manual 12 Sensor Configuration Page 26 of 50 167300IM Issue 12 Issue Date 14 10 2014...
Страница 48: ...7200 Monitor Instruction Manual 167300IM Issue 12 Issue Date 14 10 2014 Page 48 of 50...
Страница 49: ...7200 Monitor Instruction Manual 167300IM Issue 12 Issue Date 14 10 2014 Page 49 of 50...