7200 Monitor Instruction Manual
11.4 Fault Output Mode
When any of the fault conditions occur, the output can be configured to go to one of three values, two of which can be
used to remotely detect the instrument fault.
Last Valid Measurement: The output will be held at the last value that was measured before the fault occurred and will
remain held until the fault is cleared.
High/Low Value: During a fault condition the output will go to its high or low value as indicated in the table above for
the selected output type.
On entering the fault condition output menu the present setting is indicated by a tick. Use the
keys to move the
select bar to the required fault output. Press the
OK
key to make the selection, a tick will appear against the selection.
Press the menu key to exit the output type menu.
11.5 Fault Output Setup
The fault output setup determines which faults will cause the analogue output to enter its fault mode. If a fault occurs
that is not enabled, the output will remain at its last value until that fault has cleared.
The faults that can be signalled by the analogue output are displayed as a list; the enabled faults are indicated by
small ticks. To enable a fault, use the
keys to highlight the required fault and press the
OK
key. A small tick will
appear next to the selected fault. To disable a fault, use the
keys to select it and press the
OK
key, the small tick
will disappear.
Note: The instrument faults are defined elsewhere within this manual.
11.6 Test Output
This facility enables the analogue output to be tested to enable verification of its operation and also to check that
equipment connected to it is working properly. By using the
keys the output signal can be increased/decreased
over its working range in fixed steps. Press the
OK
or
MENU
to exit.
11.7 Output Test Cycle
The output test cycle can be used when installing and configuring remote equipment on the analogue output. The
analogue output is automatically swept from its low to high value enabling the operator to view the result on the remote
equipment. The output is swept for fifteen minutes after which the instrument will return to the measure mode. To exit
the output test cycle, press the
OK
or
MENU
key.
167300IM-Issue 12 Issue Date 14/10/2014
Page 25 of 50
Содержание 7200
Страница 1: ...INSTRUCTION MANUAL 7200 Monitor Field Mount...
Страница 2: ...This page is intentionally left blank...
Страница 26: ...7200 Monitor Instruction Manual 12 Sensor Configuration Page 26 of 50 167300IM Issue 12 Issue Date 14 10 2014...
Страница 48: ...7200 Monitor Instruction Manual 167300IM Issue 12 Issue Date 14 10 2014 Page 48 of 50...
Страница 49: ...7200 Monitor Instruction Manual 167300IM Issue 12 Issue Date 14 10 2014 Page 49 of 50...