143
4-2-13 Replacement of fuse
Replace fuse of the controller when blowing.
The power of the laser marker cannot be turned on if the fuse is blown.
Models of replacement parts
Be sure to use the following type of fuse: If the user applies any other fuses than the specified one, failure might result.
Laser marker model
Type
Model
LP-S500, LP-S502, LP-S505
LP-S500W, LP-S505W
Time Lag Fuse (Rating 250V 10A)
0215010
(made by Littelfuse)
LP-S200, LP-S202, LP-S205
Time Lag Fuse (Rating 250V 6.3A)
021506.3
(made by Littelfuse)
Steps for replacement
1.
Remove causes for fuse blowing.
2.
Turn OFF the key switch of the controller, and disconnect the AC power cable.
3.
Turn the fuse holder on the back side, and remove
the fuse.
4.
Replace it with new one.
5.
Turn the fuse holder, and install the fuse.
Fuse
holder
ME-LPS-SSM-8