42
Chapter 2
Cyclic Trigger Mode
In the cyclic trigger mode, the trigger is generated periodically and held in the
trigger interval.
The trigger is generated for each period.
The interval from the trigger point to the next trigger point is the sampling pe-
riod.
The analog output or judgment output is updated after the sampling period.
The TIMING indicator (green) lights up with a trigger.
In the cyclic trigger mode, no strobe output is issued.
The trigger delay time and sampling time settings are ignored.
Enter a period other than "0" for the cyclic trigger.
Example of bottom hold
Displayed value
0
Timing indicator
Internal sampling
Start of holding mode
Sampling period 1
Sampling period 3
Output of result of
sampling period 1
Output of result of
sampling period 2
ON
OFF
Sampling period 2
Trigger period
Trigger period
Trigger period
Displayed value
Output value
Measurement
Measured value
Time