52
Chapter 3
3.4 Setting According to Application (Application mode)
3.4.1 Selecting the Application Mode
In the application mode, setting items change according to each application.
Because only the items necessary for the application mode are displayed,
data entry is easy. There are the following four variations in the application
mode.
・
Manual mode: "
" (manual)
All detection functions can be used.
The manual mode is selected in the factory shipment setting.
・
Press BDC detection mode: "
" (press)
The bottom hold and trigger modes are automatically selected.
・
Rotation/eccentricity detection mode: "
" (rot)
Peak-to-peak hold is automatically selected. The minimum-based internal
trigger is selected.
Time
Displacement amount
Bottom dead center
(BDC)
Displayed value
Internal trigger level
Time
Displacement amount
Displayed value
Содержание GP-X SERIES
Страница 7: ...1 Introduction Introduction Handling Precautions 2 Checking the Packaged Components 4 Notation 4 ...
Страница 74: ...68 Chapter 3 MEMO ...
Страница 102: ...96 Chapter 4 MEMO ...
Страница 139: ...133 Chapter 7 Chapter 7 What to Do Upon an Error 7 1 Error Indication List 134 7 2 What to Do 136 ...