020-17467-00 Instructions For Use PadMixer 1000 Premium RevD
If the tubing is too large to fit through the hole, or if the tubing has large items such as y-
pieces or filters, the small drain cover plate around the drain hole can be removed.
If the mixing biocontainer incorporates an EZ Drain fitment, position it in the drain port
cutout, install the drain insert plate (see above, Figure 15), and then snap the EZ Drain clip
over the fitment OUTSIDE the tank (Figure 27 and Figure 28).
Figure 27: EZ Drain clip installation
Figure 28: EZ Drain clip installed outside the tank
Confirm that the drain is fully closed by sliding the locking collar up until it clicks, then pulling
down firmly on the blue BarbLock®* to close the drain, and sliding the locking collar down
again to lock it (Figure 29).