020-17467-00 Instructions For Use PadMixer 1000 Premium RevD
Move the system up or down to reach the desired height. The system will automatically stop
at the correct height.
Note: Always set the correct height before inflating the mixing biocontainer, in order to avoid
bag damage.
5.8
I
nflating the mixing biocontainer
Press the reset button to enable the inflation.
If the mixing biocontainer is not completely filled with liquid and the top part of the mixing
biocontainer is still empty and under vacuum the residual volume must first be filled with air
(or nitrogen).
A fully deployed mixing biocontainer is necessary to prevent the stirring mechanism from
rubbing against the side wall during mixing resulting in possible damage to the mixing
biocontainer.
Connect the inflation tubing of the mixing biocontainer to the compressed air supply and
activate the inflation (via the touch screen).
Once the mixing biocontainer reaches a certain volume the top of the mixing biocontainer will
press against the detection sensor and the compressed air supply will be automatically
stopped.
This will create an overpressure in the mixing biocontainer of +/- 10mbar. In case this is not
desired, it is not required to use the mixing biocontainer detection to stop the inflation. One
can also stop the inflation manually (by the touch screen) as soon as the mixing biocontainer
is visually sufficiently deployed as shown in Figure 41.
Figure 40: Connecting the inflation tube to the air supply
tube