020-17467-00 Instructions For Use PadMixer 1000 Premium RevD
4.3.3.8 Factory settings
Certain settings are set by the manufacturer in accordance with the type of machine.
These data are intended solely for the technician who commissions the machine and
are thus not applicable for the operator. Entering the menu of the Factory settings is not
possible with the first level password. A master password is required (5 digits). Contact
your Pall distributor for more information. Figure 19: Screen to change the Factory settings
shows these factory settings.
In the lower right hand corner of the screen one can see the current active password.
Figure 19: Screen to change the Factory settings
When pressing the
„right‟ arrow one can modify some additional parameters related to an
alarm that the system can give when during mixing the actual RPM deviates from the set
value.
One can define the time before checking the actual versus set RPM. This is set default at
60 seconds to avoid false alarms during the acceleration phase.
One can define the delay time for actually displaying the
„out of range‟ message. This is set
default at 30 seconds.
One can define the maximum RPM deviation allowed. This is set default at 20 RPM.