020-17467-00 Instructions For Use PadMixer 1000 Premium RevD
17
Figure 20: Screen to change additional Factory settings
4.3.4. Inflate
This function is used to inflate the mixing biocontainer with compressed air or nitrogen.
When activated the icon turns green and the compressed air valve is activated.
The compressed air remains active until one of the following points:
The function is cancelled by touching the INFLATE icon on the touch screen again
The mixing biocontainer is inflated to the point that it presses against the mixing
biocontainer detection. However, if the mixing biocontainer is emptied and the
detection is released the compressed air will reactivate after a delay of three
seconds
The emergency stop is pressed
The protective front cover is opened.
4.3.5. Print
This prints out the data for the last finished recipe. To print the last recipe, make sure the
mixing system has completely stopped rotating.
Note: the total mixing time stated on the label may be a few seconds shorter than the
difference between the start and finish times. The reason is that the last deceleration time is
not included in the calculation of the total mixing time, but the finish time is not
determined until the mixer has come to a complete standstill.
4.3.6. Reset
This resets all error messages, for example after using the emergency stop.