020-17467-00 Instructions For Use PadMixer 1000 Premium RevD
1. Introduction
The purpose of this manual is to inform the users about the safe and correct use of the
PadMixer 1000 Premium. This includes all persons who have the task of installing,
commissioning, adjusting, maintaining, cleaning, repairing, transporting or scrapping the
machine. Before using this machine all persons involved should thoroughly read and
understand the user manual. All safety precautions and safety provisions in use in the
company must be known and respected.
This user manual must always be available near the machine. It is very important that all
instructions are carefully followed and, where appropriate, they should be incorporated into
the user’s standard operating procedures. If some of the procedures do not suit your needs,
please consult your Pall distributor before finalizing your system. Use of this product in a
manner other than in accordance with Pall’s current recommendations may lead to injury or
product loss. Pall cannot accept liability for such injury or loss.
2. Safety warnings
Operation outside the specifications defined in this manual may cause personal injury
and result in damage to the equipment.
No operations (cleaning, maintenance, ...) may be performed on the machine if the
machine is not switched off. Protections around moving parts may not be removed.
3. Assembly and commissioning
It is not permitted to move the machine when a tank and/or mixing biocontainer is connected
to it. To move the machine push it from the backside.
Do not attempt to move the machine by pushing/pulling the mixing head.
Required facilities:
Electricity:
The system 701024 can be used on a single phase net of 220V.
The system 701039 can be used on a single phase net of 110V.
The fuse present in the local power supply cabinet has to be minimum
16Ampère, with a curve C or curve D.
In case a differential switch is present in the local power supply
cabinet, it has to be type A or type B.
Compressed air:
Min. 3 bar, max. 8 bar.
Use of nitrogen is permitted.
The connection of the system to the compressed air system is effected
by connecting a 6 mm ID compressed air hose to the compressed air
coupling closest to the rear of the machine.
A compressed air cable is provided on the coupling closest to the front
of the machine with a fast coupling to be connected to the mixing
biocontainer.
The outgoing pressure (to the mixing biocontainer) can be adjusted by
means of the manometer underneath the mixing head cover plate. This
pressure is preset to a maximum of 500 mbar.