Macro 900 Operation Manual
V3
© 2013 Palintest Ltd.
www.palintest.com
Page 88 of 147
x
Differences in the fluorescent response between various types of oil.
x
Differences in the fluorescent response caused by temperature.
x
Differences in the fluorescent response caused by ambient light.
x
Interference caused by turbidity.
Fluorescence measurement techniques are ideal for researchers who are
interested in detecting the presence or absence of a specific substance and
measuring relative fluorescence changes that can be used as an indication of
increasing or decreasing concentrations.
Fluorescence measurement techniques are not ideal for quantitative
measurement.
In order to obtain more accurate results, data obtained
with the fluorometer in the field should be post-calibrated with data
from standard laboratory analysis of grab samples acquired during the
study.
15.7.3.
Special Precautions When Using the REFOIL Probe
x
Always observe the safety advice printed above.
x
The UV LED is permanently on when power is supplied to the MAP
2000/2100 and draws 50mA. Switch the MAP 2000/2100 off between
readings to save power and prolong the UV LED's life.
x
Do not leave the MAP 2000/2100 powered on for prolonged periods if not
submerged. This can cause the UV LED to overheat.
x
Do not deploy the REFOIL probe in water temperatures above 30ºC.
x
Always carry out a zero point (Point 1) calibration as close to
operational temperature as possible prior to every use.
15.7.4.
Calibrating the REFOIL Probe
The REFOIL probe has two calibration points. Careful calibration is essential in
order to ensure consistent and reliable results.
When a REFOIL probe is first
installed,
it MUST be calibrated at both points
in order to set the probe's
relative sensitivity and establish its slope.
Subsequently, a single point (Zero) calibration should be carried out daily.
Calibration of the REFOIL probe Zero point is normally carried out during
MacroCal (see
Full two-point calibration should be carried out every few months.
15.7.5.
Calibration Solution Preparation
In order to 'calibrate' the REFOIL probe, a 10ppm calibration solution of 1-5,
naphthalenedisulfonic acid disodium salt should be used. This solution contains
naphthalene, an aromatic hydrocarbon, which has
similar
fluorescence
characteristics to many Refined Oils.
The 10ppm calibration solution should be freshly prepared by serial dilution
from pure 1-5, naphthalenedisulfonic acid disodium salt. The following
Naphthalene salt is recommended:
Содержание Macro 900
Страница 1: ...Macro 900 Water Quality System Operation Manual www palintest com ...
Страница 77: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 77 of 147 ...
Страница 92: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 92 of 147 ...
Страница 132: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 132 of 147 2004 108 EC and 1999 5 EC ...