temperature in +/-millivolts (mV). During normal operation this offset will be
subtracted from the ORP electrode output to give a corrected ORP display.
This value is not stored in memory so should be noted down in a calibration
record book for the probe. When calibration is complete, press the
OK
key then
the
ESC
key repeatedly to return to normal operating mode.
Remove the Probe from the calibration bottle, rinse thoroughly in fresh water,
shake off any excess and dry the outer sleeve with a soft cloth. Dampen the
sponge in the storage cap with storage solution and fit it to the pH/ORP
electrode. ORP calibration is now complete.