Calibrating
100%
Press [OK]
When calibration is complete, press
OK
then
ESC
to return to normal reading
mode
.
Important
If you have both Optical and ISE electrodes fitted to your Macro Accessory
Probe, you must now remove the rubber sealing cap from the ISE electrode
then re-calibrate the optical probe's zero point (Pt-1) in fresh, clean water
(bottled still mineral water is ideal).
The ISE electrode's sealing cap alters the optical characteristics of the
measurement chamber and can therefore render the optical electrode's zero
point calibration performed during MacroCal erroneous.
Now the DO 100% saturation point should be calibrated in damp air.