Macro 900 Operation Manual
V3
© 2013 Palintest Ltd.
www.palintest.com
Page 145 of 147
26.
Appendix 6. ISE Electrodes Detailed Specification
Range
0
–
9,000mg/L (ppm)
Resolution
2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0
–
8,999.9 mg/L
Ammonium /
Accuracy
± 10% of reading or 2ppm (whichever is greater)
Ammonia
†
MLD
(1)
1.0 ppm
Interfering
Ions
(2)
Potassium, Sodium and Magnesium
pH Range
(3)
5 - 8
Range
0
–
20,000mg/L (ppm)
Resolution
2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0
–
19,999.9 mg/L
Chloride
Accuracy
± 10% of reading or 2ppm (whichever is greater)
MLD
(1)
2.0 ppm
Interfering
Ions
(2)
Bromide, Iodide, Cyanide and Sulphide
pH Range
(3)
2 - 11
Range
0
–
1,000mg/L (ppm)
Resolution
2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0
–
999.9 mg/L
Fluoride
Accuracy
± 10% of reading or 2ppm (whichever is greater)
MLD
(1)
0.05 ppm
Interfering
Ions
(2)
Hydroxide (OH-)
pH Range
(3)
4 - 8
Range
0
–
30,000mg/L (ppm)
Resolution
2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0
–
29,999.9 mg/L
Nitrate
Accuracy
± 10% of reading or 2ppm (whichever is greater)
MLD
(1)
0.5 ppm
Interfering
Ions
(2)
Chloride, Bromide, Fluoride, Sulphate, Chlorate and Perchlorate
pH Range
(3)
3 - 10
Range
0
–
2,000mg/L (ppm)
Resolution
2 Auto-range scales: 0.00 - 99.99 mg/L, 100.0
–
1,999.9 mg/L
Calcium
Accuracy
± 10% of reading or 2ppm (whichever is greater)
MLD
(1)
0.05 ppm
Interfering
Ions
(2)
Magnesium, Barium, Lead, Zinc and Sodium
pH Range
(3)
4 - 9
†
Ammonia readings are calculated from Ammonium, pH and temperature electrode values.
Palintest Ltd. reserves the right to change specifications without notice
Notes:
1.
MLD (Minimum Level of Detection) is the minimum value the electrode is
physically capable of measuring.
2.
Each ion selective electrode is prone to interference from ions that are
similar in nature to the target ion. The main interfering ions for each
electrode type are listed here. If the water under test contains interfering
ions, the electrode will produce erroneous readings.
Ion Selective
Electrodes are not recommended for use in brackish or salt water
due to the high level of interfering ions.
Содержание Macro 900
Страница 1: ...Macro 900 Water Quality System Operation Manual www palintest com ...
Страница 77: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 77 of 147 ...
Страница 92: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 92 of 147 ...
Страница 132: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 132 of 147 2004 108 EC and 1999 5 EC ...