15.3.6.
Zero Point Calibration
To calibrate the zero point, follow these steps:
1.
Fill a calibration bottle with distilled water, remove the storage cap from
the pH probe if fitted, wash the Probe in distilled water, then drop the
Probe in all the way.
The Sleeve End Cap and Plug must be fitted
.
Bang the Probe against the bottom of the bottle several times in order to
remove any air bubbles that may be clinging to the probe.
2.
Switch the Macro 900 Meter on and wait until the temperature and BGA-
PE readings are stable. If the BGA-PE reading is very high, there are
probably air bubbles adhering to the lenses. Bang the Probe against the
bottom of the bottle to remove.
3.
Ensure the temperature of the solution is between 5ºC and 40ºC (41ºF -
104ºF).
4.
Press the
MENU
key then select
Calibration.
The following screen will be
displayed.
Calibration
Æ
MacroCal
DO 100%
Full Cal
5.
Select
Full Cal.
The screen will change to:
Calibration
Æ
pH/ORP
DO/EC
Aux Probes
6.
Select
Aux Probes
. The screen will change to:
SELECT PROBE
Æ
1:BGA-PE l 4:N/A
2:EMPTY l 5:N/A
3:N/A l 6:N/A
The BGA-PE probe should have been assigned to AUX socket 1
when it was fitted. Press the OK or right arrow key to select BGA-
PE. The screen will change to:
Содержание Macro 900
Страница 1: ...Macro 900 Water Quality System Operation Manual www palintest com ...
Страница 77: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 77 of 147 ...
Страница 92: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 92 of 147 ...
Страница 132: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 132 of 147 2004 108 EC and 1999 5 EC ...