The Calibration Report on the top line displays the EC Cell Constant. This value
is not stored in memory so should be noted down in a calibration record book
for the probe.
When the cell constant details have been noted down, press
OK
then
ESC
repeatedly to return to normal reading mode
.
Special Notes:
¾
If you have selected a Calibration Standard value other than 2570
(MacroCal), then you subsequently use the MacroCal calibration
technique described in section 10, the Calibration Standard value
will automatically be reset to 2570.
¾
The Calibration Standard value is stored in the Probe,
not
the
Meter. If you use one Meter with several different Probes, you will
have to set the Calibration Standard value for each probe
individually during calibration.
¾
If you select a Calibration Standard value but do not press
OK
, the
information will not be sent to the Probe and the change will not be
registered.
14.9.
Verifying EC Calibration
Due to the fact that debris and air bubbles can adversely affect EC calibration,
it is advisable to verify calibration has been properly achieved. To do this,
follow item nine above with this procedure.
1.
Remove the probe from the calibration bottle, shake it off, then
reinsert.
2.
Press the
ESC
key repeatedly to get back to the Main Menu.
3.
Go into settings and make sure EC is set to read with reference to
25ºC. If it’s
not, set it that way. See section 9
Error! Reference
source not found.
.
4.
Go back to the main screen, wait until the temperature and EC
readings are
stable, then check that the EC is r/- 1%
of the Calibration Standard
value.
5.
If the EC reading is outside the 1% limit, recalibrate, this time
leaving more
time for stabilisation.
If you cannot successfully verify the EC calibration after several attempts,
replace the Calibration Standard. If the problem persists, strip the probe down
as described below and thoroughly clean the EC contacts.
14.10.
Errors During Calibration
At the beginning of the calibration routine, a sanity check is done. If the probe
detects that the Calibration Standard value set and the Calibration Standard
being used differ, the 'OUT OF CAL RANGE' error will be reported. If any other
problems occur during calibration, an error message will be displayed. Refer to
Error! Reference source not found.
in section 10 for error message
handling.
Содержание Macro 900
Страница 1: ...Macro 900 Water Quality System Operation Manual www palintest com ...
Страница 77: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 77 of 147 ...
Страница 92: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 92 of 147 ...
Страница 132: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 132 of 147 2004 108 EC and 1999 5 EC ...