
Macro 900 Operation Manual
V3
© 2013 Palintest Ltd.
www.palintest.com
Page 105 of 147
16.6.2.
Three-point Calibration
During three-point calibration, the MAP 2000/2100 and Macro 900 Meter must
remain switched on. If the Macro 900 Meter is switched off between points, the
calibration process will be aborted and must be re-started from point 1. To
calibrate the ISE electrode follow these steps:
Point 1.
1.
Remove the storage cap from the pH electrode, wash the Probe in
distilled water, dry the probe thoroughly then drop the Probe in to the
warm
10ppm
solution.
2.
Switch the Macro 900 Meter on and stir the probe until the temperature
and Cl readings are completely stable. A minimum of five minutes is
recommended.
3.
Ensure the temperature of the solution is between 20ºC and 40ºC (68ºF -
104ºF).
4.
Press the
MENU
key then select
Calibration.
The following screen will be
displayed.
Calibration
Æ
MacroCal
DO 100%
Full Cal
5. Select
Full Cal.
The screen will change to:
Calibration
Æ
pH/ORP
DO/EC
Aux Electrodes
6. Select
Aux Electrodes
. The screen will change to:
SELECT ELECTRODE
Æ
1:TURB l 4:N/A
2:Cl l 5:N/A
3:N/A l 6:N/A
The Chloride (Cl) electrode should have been assigned to AUX socket 2
when it was fitted. Move the pointer down to 2:Cl then press the OK or
right arrow key to select.
7. The screen will change to:
CALIBRATE Cl
Æ
Pt-1? [01/Jan/12]
Pt-2? [01/Jan/12]
Pt-3? [01/Jan/12]
Содержание Macro 900
Страница 1: ...Macro 900 Water Quality System Operation Manual www palintest com ...
Страница 77: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 77 of 147 ...
Страница 92: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 92 of 147 ...
Страница 132: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 132 of 147 2004 108 EC and 1999 5 EC ...