Macro 900 Operation Manual
V3
© 2013 Palintest Ltd.
www.palintest.com
Page 49 of 147
7.
Select
DO/EC
. The screen will change to:
Calibration
Æ
DOZero?[01/Jan/12]
DO100%?[01/Jan/12]
EC2570?[01/Jan/12]
The dates shown to the right of the screen are the dates of the last
successful calibration. The value shown on the bottom line next to
‘EC’ is the value the EC
electrode was last calibrated to.
8.
Move the pointer down to the bottom line using the down arrow
key.
Calibration
DOZero? [01/Jan/12]
DO100%?[01/Jan/12]
Æ
EC2570?[01/Jan/12]
If the Calibration Standard value you are using is already displayed,
press the
OK
key to start calibrating. Remember, if you are using
MacroCal solution, the EC value on this line should be 2570.
If the value of the EC Calibration Standard you are using is not
displayed, press the right arrow key. The bottom line will change to:
Calibration
DOZero?[01/Jan/12]
DO100%?[01/Jan/12]
EC
Æ
2570?[01/Jan/12]
You can now use the up and down arrow keys to select one of three
EC Calibration Standard values (1413, 2570 or 12880).
9.
Once the correct Calibration Standard value is being displayed,
press the
OK
key. The screen will change to:
PLEASE WAIT
Stabilising
000%
10.
If the calibration is successful, the counter will reach 100% and the
following screen will be displayed.
Cell constant:0.98
Calibrating
100%
Press [OK]
Содержание Macro 900
Страница 1: ...Macro 900 Water Quality System Operation Manual www palintest com ...
Страница 77: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 77 of 147 ...
Страница 92: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 92 of 147 ...
Страница 132: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 132 of 147 2004 108 EC and 1999 5 EC ...