Macro 900 Operation Manual
V3
© 2013 Palintest Ltd.
www.palintest.com
Page 101 of 147
Calibration point 1 (Pt-1) is the warm 10ppm point. Calibration point 2
(Pt-2) is the warm 100ppm point. Calibration point 3 (Pt-3) is the cool
10ppm point.
The dates shown to the right of each point are the dates of the last
successful calibration.
8. Select Pt-1. The screen will change to:
PLEASE WAIT
Stabilising
000%
The Meter will wait until the readings are stable, then it will send the calibration
command to the Probe, where the calibration takes place. During calibration,
the Calibrating screen is displayed and the progress counter counts up. If the
calibration is successful, the counter will reach 100% and the following screen
will be displayed.
Output:348mV
Calibrating
100%
Press [OK]
The Calibration Report on the top line displays the voltage output from the ISE
electrode in millivolts (mV). This value is not stored in memory so should be
noted down in a calibration record book for the probe along with the
temperature.
Point 2
1.
Remove the probe from the 10ppm solution and wash thoroughly in
deionised water. Dry the probe then drop it into to the warm
100ppm
solution.
2.
Stir the probe until the temperature and NO3 readings are completely
stable. A minimum of five minutes is recommended.
3.
Ensure the temperature of the solution is within 1ºC of the
previous 10ppm calibration point.
If the solution is warmer or cooler
than this, calibration will fail.
4.
Referring to steps 4-7 above, select Pt-2 and press OK.
The Meter will wait until the readings are stable, then it will send the calibration
command to the Probe, where the calibration takes place. During calibration,
the Calibrating screen is displayed and the progress counter counts up as
shown above. If the calibration is successful, the counter will reach 100% and
the calibration report screen will be displayed.
If the temperature of the 100ppm solution is more than 1ºC different from the
Содержание Macro 900
Страница 1: ...Macro 900 Water Quality System Operation Manual www palintest com ...
Страница 77: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 77 of 147 ...
Страница 92: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 92 of 147 ...
Страница 132: ...Macro 900 Operation Manual V3 2013 Palintest Ltd www palintest com Page 132 of 147 2004 108 EC and 1999 5 EC ...