Macro 900 Operation Manual
V3
© 2013 Palintest Ltd.
www.palintest.com
Page 41 of 147
4.
Press the
MENU
key then select
Calibration.
The following screen
will be displayed.
Calibration
Æ
MacroCal
DO 100%
Full Cal
5.
Select
Full Cal.
The screen will change to:
Calibration
Æ
pH/ORP
DO/EC
Aux Electrodes
6.
Select
pH/ORP
. The screen will change to:
Calibration
Æ
pH7.00?[01/Jan/12]
pH4.01?[01/Jan/12]
ORP? [01/Jan/12]
7.
Select ORP. The screen will change to:
PLEASE WAIT
Stabilising
000%
The Meter will wait until the readings are stable, then it will send the calibration
command to the Probe, where the calibration takes place. During calibration,
the Calibrating screen is displayed and the progress counter counts up. If the
calibration is successful, the counter will reach 100% and the following screen
will be displayed.
Offset: 5.5mV
Calibrating
100%
Press [OK]
The Calibration Report on the top line displays the voltage offset between the
ORP electrode output and the value of the calibration solution at the calibration