25
English
Power, weight & size:
Rated power: 12V, type
Instrument size: 203mm x 163mm x 72mm
Package size: 270mm x 243mm x 105mm
Instrument weight: 0.85kg
Gross weight: 2.1kg
Environmental conditions during transport:
Temperature: -40ºC to +70ºC
Relative humidity: 10% to 95%
Pressure: 500 hPa to 1060 hPa
Environmental conditions during use:
Temperature: +10ºC to +40ºC
Relative humidity: 30% to 75%
Pressure: 700 hPa to 1060 hPa
IP class: IP20
17) Troubleshooting
Problem
Cause
Possible Solution
No measurement
Probe not connected.
Connect the probe.
Electromagnetic interference.
Remove the source of electromagnetic
interference.
Probe is held too far away from
the SmartPeg.
Hold the probe 3-5mm away from the
SmartPeg.
Charger does not fit
the electrical outlets
Wrong Osstell plug adapter.
Use correct Osstell plug adapter.
Instrument does not
start
Uncharged battery.
Charge the Osstell IDx.
Difficult to measure in
an exact buccolingual
direction
No space.
Try to measure at a slightly different rotational
angle.
Difficulties screwing in
the SmartPeg
Wrong SmartPeg.
Ensure that the SmartPeg is compatible with
the implant system.
The instrument does
not work
The Osstell IDx has not been
registered.
Register the instrument at
OsstellConnect.com
or contact service.
The probe is not connected
properly.
Connect the probe.
The tooth view is gone
Classic examination mode is
activated.
Deactivate classic mode, see SETTINGS
section.
If your problem continues, please contact your distributor or our service team for further assistance.
In the event of an instrument malfunction, the Osstell IDx and accessories should be sent to the manufacturer
for repair.
18) Service
Osstell AB
Phone: +46 31 340 82 50
Stampgatan 14
Fax: +46 31 413 115
411 01 Göteborg, Sweden
Email: [email protected]
Содержание IDx