DSPEC Pro
®
Digital Gamma-Ray Spectrometer User’s Manual
794380D / 0914
62
LIST_MODE_ZDT
Reports the legal ZDT-mode settings. Each item in the list can be selected by sending the
SET_MODE_ZDT command with a mode number. The mode number is implied by the loca-
tion of the mode in the list. For example, the first item in the list had a mode number of 0. To
select the first item, send SET_MODE_ZDT 0. The second item has a mode number of 1,
and so on.
Response:
MODE_ZDT NORM_CORR CORR_ERR
Mode 0 stores the normal and corrected
spectra. Mode 1 stores the corrected and
the error spectra.
LIST_ROI_SAMPLE
Lists the valid ROI sample settings.
Response:
ROI_SAMP PUR BLN BLRG BLD BUSY GATE RESV PKDET
LIST_SHAP_CUSP
Not used by the DSPEC Pro; included for backward compatibility.
LIST_SHAP_FLAT
Lists the valid flattop width settings.
Response:
SHAP_FLAT 0.3 2.4 3 24
Settings are 0.3 to 2.4 in 22 steps of
0.1
µ
s.
LIST_SHAPE_RISE
Lists the valid rise-time settings.
Response:
SHAP_RISE 0.8 23.0 4 115
Settings are 0.8 to 23 in 0.2-
µ
s steps.
LIST_SHUT
Lists the valid shutdown settings.
Response:
SHUTDOWN TTL ORTEC SMART
LIST_TIME_ETP
Returns the range of valid protection-time settings in the Enhanced Throughput Mode.
Response:
TIME_ETP 01 200 4 8000
Valid settings range from 0.1
µ
s to 200
µ
s
in 25-ns increments.