794380D / 0914
2. GETTING STARTED
41
Figure 24. DSPEC Pro MDA Preset Tab.
where:
a, b, and c are determined by the MDA criteria you choose.
Counts is the gross counts in an ROI that is 2.5×FWHM around the target peak energy.
Live time is evaluated in 40 second intervals for the MDA presets.
CorrectionFactor is the product of the calibration efficiency at the specified peak energy and
the peak’s branching ratio (yield) as listed in the working (active) library.
NOTE
MAESTRO does not support efficiency calibration. The efficiency component in
the CorrectionFactor is set to 1.0; the preset field is labeled Correction instead of
MDA; and the preset is based on counting activity (cA) instead of becquerels. You
can enter the MDA preset either in counts; or corrected for factors such as sample
volume, attenuation, or calculated efficiency. For example, if you manually calcu-
late the efficiency for a peak, you can enter a corrected MDA target value by multi-
plying the desired MDA value times the calculated efficiency, and entering the
product as the Correction.