Figure 15. DSPEC Pro InSight Mode.
The Status Sidebar changes from the PHA mode controls to the InSight controls for adjusting
the peak display. On the left is a vertical scrollbar for adjusting the vertical offset of the wave-
form. The value of the offset is shown on the display. Double-clicking the mouse in the scrollbar
will set the vertical offset to the vertical value of the channel at the marker position. This lets
you conveniently zoom in on a particular part of the waveform (such as the tail for pole-zero-
ing).
In the Auto trigger mode, the display is updated every time a new pulse exceeds the trigger
level. To keep a single pulse displayed, select Single. Click Reset to refresh the display to see
the next pulse. There will usually be one or two pulses in the “pipeline” that will be displayed
before any change entered will be seen. If the trigger is turned off, the display will be redrawn
periodically, even if no pulse is there.
The Delay setting is the time delay between the pulse shown on the display and the trigger level
crossing. The value of the time delay is shown on the display.