4. FIRMWARE COMMANDS AND RESPONSES
65
SET_GAIN_CONVERSION 2048<CR>
Conversion gain set to 2048 channels.
SET_GAIN_CONVERSION 4096<CR>
Conversion gain set to 4096 channels.
SET_GAIN_CONVERSION 8192<CR>
Conversion gain set to 8192 channels.
SET_GAIN_CONVERSION 16384<CR>
Conversion gain set to 16384 channels.
SET_GAIN_FINE value
Sets the fine gain to value. Value is a floating point value from 0.333 to 1.0. See also
SHOW_GAIN_FINE.
SET_GAIN_NEGATIVE
Sets the amplifier input polarity to negative.
SET_GAIN_POSITIVE
Sets the amplifier input polarity to positive.
SET_GAIN_PRESET count
No function in the DSPEC Pro; included for backward compatibility.
SET_GAIN_WIDTH chans
Sets the width in channels for the stabilizer gain peak. The gain width must be chosen such
that the beginning channel is no lower than channel 0 and the ending channel is no higher
than the maximum channel as determined by the conversion gain. The gain channel and
width must be set before gain stabilization can be enabled. The absolute minimum width for
the gain peak is 3 channels, and the absolute maximum width for the gain peak is 256 chan-
nels in Gauss mode. In point mode there is no maximum peak width, though the chosen
width must allow the peak to fit within the device’s channel limits as stated above. See also
SHOW_GAIN_WIDTH, SET_GAIN_CHANNEL and SHOW_GAIN_CHANNEL.
SET_GATE_ANTICOINCIDENT
Causes the DSPEC Pro to use the ADC gate input signal in anticoincident mode. See also
SET_GATE_OFF, SET_GATE_COINCIDENT, and SHOW_GATE.
SET_GATE_COINCIDENT
Causes the DSPEC Pro to use the ADC gate input signal in coincident mode. See also
SET_GATE_OFF, SET_GATE_ANTICOINCIDENT, and SHOW_GATE.
SET_GATE_OFF
Causes the DSPEC Pro to ignore the state of the ADC gate input signal. See also
SET_GATE_COINCIDENT, SET_GATE_ANTICOINCIDENT, and SHOW_GATE.