background image

23

NOISE

Measurethenoiseat the amplifier Unipolar

output with maximum amplifier gain and 2 µs

shaping time. Using a true rms voltmeter, the noise

should be less than 5 µV x 1500 (gain), or 7.5 mV.
For an average responding voltmeter, the noise

reading would have to be multiplied by 1.13 to

calculate the rms noise. The input must be

terminatedin100

duringthenoisemeasurements

6.3. SUGGESTIONS FOR

TROUBLESHOOTING

In situations where the 572A is suspected of a

malfunction, it is essential to verify such

malfunction in terms of simple pulse generator

impulses at the input. The 572A must be

disconnected from its position in any system, and

routine diagnostic analysis performed with a test

pulse generator and an oscilloscope. It is

imperative that testing not be performed with a

source and detector until the amplifier performs

satisfactorily with the test pulse generator.
The testing instructions in Section 6.2 and the

circuit descriptions in Section 5 should provide

assistance in locating the region of trouble and

repairing the malfunction. The two side plates can

be completely removed from the module to enable

oscilloscope and voltmeter observations.

6.4. FACTORY REPAIR

This instrument can be returned to the ORTEC

factory for service and repair at a nominal cost. Our

standard procedure for repair ensures the same

quality control and checkout that are used for a

newinstrument.AlwayscontactCustomerServices

at ORTEC, (865) 482-4411, before sending in an

instrument for repair to obtain shipping instructions

and so that the required Return Authorization

Number can be assigned to the unit. This number

should be marked on the address label and on the

package to ensure prompt attention when the unit

reaches the factory.

6.5. TABULATED TEST POINT VOLTAGES

The voltages given in Table 6.1 are intended to

indicate typical dc levels that can be measured on

the printed circuit board. In some cases the circuit

will perform satisfactorily even though, due to

component tolerances, there may be some voltage

measurements that differ slightly from the listed

values. Therefore the tabulated values should not

be interpreted as absolute voltages but are

intended to serve as an aid in troubleshooting.

Table 6.1. Typical dc Voltages

Note: All voltages measured with no input signal,

with the input terminated in 100Ω, and all controls

set fully clockwise at maximum.

Location

Voltage

TP1

±5 mV

TP2

±30 mV

TP3

±20 mV

TP4

±20 mV

TP5

±30 mV

TP6

0 to +3.3 V

TP7

±6 mV

Q15E

-15 V ±0.8 V

Q16E

+15 V ±0.8 V

IC13 pin 2

+5 V ±0.3 V

Содержание 572A

Страница 1: ...Model 572A Spectroscopy Amplifier Operating and Service Manual Printed in U S A ORTEC Part No 785100 0904 Manual Revision D...

Страница 2: ...nbeassignedtotheunit Also ORTECmust be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the fault of the instrument being returned and of...

Страница 3: ...NS 7 3 7 SHORTING OR OVERLOADING THE AMPLIFIER OUTPUTS 8 3 8 INHIBIT OUTPUT CONNECTION 8 3 9 BUSY OUTPUT CONNECTION 8 3 10 CRM OUTPUT CONNECTION 8 4 OPERATING INSTRUCTIONS 8 4 1 INITIAL TESTING AND OB...

Страница 4: ...a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please re...

Страница 5: ...he instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug...

Страница 6: ...vi...

Страница 7: ...m 100 mV to 100mV Thisoutputpermitstheuse of the direct coupled input of the analyzer with a minimum amount of interface problems The 572A bipolar output may be preferable for spectroscopy when operat...

Страница 8: ...seline with the preamplifier time constant The pole zero adjust is accessible from the front panel of the 572A and can easily be adjusted by observing the baselinewithanoscilloscopewithamonoenergetic...

Страница 9: ...3 Fig 1 1 Differentiation in an Amplifier Without Pole Zero Cancellation Fig 1 2 Differentiation in a Pole Zero Canceled Amplifier...

Страница 10: ...o 5OEC CROSSOVER WALK 3 ns for 50 1 dynamic range includingcontributionofORTEC551or552 Constant FractionTimingSingle ChannelAnalyzer using 50 fraction and 0 5 s shaping COUNT RATE STABILITY The1 33MeV...

Страница 11: ...iometeradjuststheunipolar output baseline dc level range 100 mV to 100 mV 2 3 INPUT INPUT Type BNC front and rear panel connectors accept either positive or negative pulses with rise times in the rang...

Страница 12: ...e pole zero adjustment see Section 4 6 An alternatemethodisaccomplishedeasilybyusinga monoenergetic source and observing the amplifier baseline with an oscilloscope after each pulse under approximatel...

Страница 13: ...h shaping time constant is selected The 572A provides both unipolar and bipolar outputs Theunipolaroutputpulsesshouldbeused in applications where the best signal to noise ratio resolution is most impo...

Страница 14: ...e pileup is sensed in the 572A and the pulse can then be used to prevent the Analyzer from measuring and storing a false amplitude The signal is dc coupled and rises from 0toabout 5Vforatimeequalto6 s...

Страница 15: ...lates the threshold of the BLR gate according to the output noise level The Threshold setting permits manual control of the BLR gate threshold usingthescrewdrivercontrolimmediatelybelowthe toggle swit...

Страница 16: ...on power in the Bin and Power Supply and allow the electronics of the system to warm up and stabilize b Set the 572A controls initially as follows Shaping 2 s Coarse Gain 50 Gain 1 000 Internal Jumper...

Страница 17: ...of the preamplifier Adjusttheattenuatorsothatthe572A output amplitude is about 9 V d Observe the unipolar output of the 572A with an oscilloscope triggered from the 572ABusyoutput Adjust the PZ Adj co...

Страница 18: ...tal deflection Trigger the oscilloscope with the Busy output from the 572A d Reduce the control setting until the baseline discriminator begins to trigger on noise this corresponds to about 200 counts...

Страница 19: ...r for the same pulse height as the pulse obtained in stepc LocktheNormalizecontrolanddonotmove it again until recalibration is required The pulser is now calibrated the Pulse Height dial read directly...

Страница 20: ...capacitance decreases thus reducing the resolution spread The overall resolution spread will depend upon which effect is dominant Figure4 7showscurvesoftypicalnoise resolution spread versus bias volt...

Страница 21: ...PECTROSCOPY SYSTEMS HIGH RESOLUTION ALPHA PARTICLE SPECTROSCOPYSYSTEM Theblockdiagramof a high resolution Spectroscopy system for measuringnaturalalphaparticleradiationisshown in Fig 4 11 Alpha partic...

Страница 22: ...ounters operated at high gas amplification tend to degrade the resolution capabilities drastically at x ray energies even at relatively low counting rates By using a high gain low noise amplifying sys...

Страница 23: ...17 Fig 4 13 Scintillation Counter Gamma Spectroscopy System Fig 4 14 High Resolution X Ray Energy Analysis System Using a Proportional Counter Fig 4 15 General System Arrangement for Gating Control...

Страница 24: ...18 Fig 4 16 Gamma Ray Charged Particle Coincidence Experiment Fig 4 17 Gamma Ray Pair Spectrometry...

Страница 25: ...ant of 0 5 1 2 3 6 or 10 s The amplifier produces the fastshapedpulsefromthesamepreamplifierinput pulse and this triggers discriminator IC12 set just above the noise level The discriminator response t...

Страница 26: ...20 Fig 5 1 Amplifier Block Diagram...

Страница 27: ...ackard 3400A RMS Voltmeter 6 2 PULSER TEST1 Coarse Gain 1K Gain 1 5 Input Polarity Positive Shaping Time Constant 1 s BLR PZ Adj Variable control Fully CW for 300 mV a Connect a positive pulser output...

Страница 28: ...y to vary the PZ Adj control on the front panel in order to cancel the pulser pole and minimize the time required for return to the baseline Increasethepulseroutputamplitudeto1000times the setting tha...

Страница 29: ...rvations 6 4 FACTORY REPAIR This instrument can be returned to the ORTEC factoryforserviceandrepairatanominalcost Our standard procedure for repair ensures the same quality control and checkout that a...

Страница 30: ...served bus 34 Power return ground 13 Spare 35 Reset Scaler 14 Spare 36 Gate 15 Reserved 37 Reset Auxiliary 16 12 V 38 Coaxial 17 12 V 39 Coaxial 18 Spare bus 40 Coaxial 19 Reserved bus 41 117 V ac neu...

Отзывы: