background image

7

direct connection. When any one of these units is

used, it should be terminated with a 100Ω

terminator at the amplifier input or be used with at

least one of the output attenuators set at In. (The

small error due to the finite input impedance of the

amplifier can normally be neglected.)

SPECIAL CONSIDERATIONS FOR POLE-ZERO

CANCELLATION

When a tail pulser is connected

directly to the amplifier input, the PZ Adj should be

adjusted if overload tests are to be made (other

tests are not affected). SeeSection4.6forthe pole-

zero adjustment. If a preamplifier is used and a tail

pulser is connected to the preamplifier test input,

similar precautions are necessary. In this case the

effect of the pulser decay must be removed; i.e., a

step input should be simulated.

3.5. SHAPING CONSIDERATIONS

The shaping time constant on the 572A is switch-

selectable in steps of 0.5, 1, 2, 3, 6, and 10 µs. The

choice of the proper shaping time constant is

generally a compromise between operating at a

shorter time constant for accommodation of high

counting rates and operating with a longer time

constant for a better signal-to-noise ratio. For

scintillationcounterstheenergyresolutiondepends

largely on the scintillator and photomultiplier, and

therefore a shaping time constant of about four

times the decay-time constant of the scintillator is a

reasonable choice (for Nal, a 1-µs shaping time

constant is about optimum). For gas proportional

counters the collection time is normally in the 0.5 to

5 µs range and a 2 µs or greater time constant

selectionwillgenerallygiveoptimumresolution.For

surface barrier semiconductor detectors, a 0.5 to 2

µs resolving time will generally provide optimum

resolution. Shaping time for Ge(Li) detectors will

vary from 1 to 6 µs, depending on the size,

configuration, and collection time of the specific

detectorandpreamplifier.Whenacharge-sensitive

preamplifier is used, the optimum shaping time

constant to minimize the noise of a system can be

determined by measuring the output noise of the

systemand dividing it by the system gain.Sincethe

572A has almost constant gain for all shaping

modes,theoptimumshapingcanbedeterminedby

measuring the output noise of the 572A with a

voltmeter as each shaping time constant is

selected.
The 572A provides both unipolar and bipolar

outputs. The unipolar outputpulsesshould be used

in applications where the best signal-to-noise ratio

(resolution) is most important, such as high-

resolution spectroscopy using semiconductor

detectors. Use of the unipolar output with baseline

restorationwillalsogiveexcellentresolutionathigh

counting rates. The bipolar output should be used

in high count rate systems when the analyzer

system is ac-coupled and noise, or resolution, is a

secondary consideration.

3.6. LINEAR OUTPUT CONNECTIONS

AND TERMINATING CONSIDERATIONS

Since the 572A unipolar output is normally used for

spectroscopy the 572A is designed with a great

amount of flexibility in order for the pulse to be

interfaced with an analyzer. A gated baseline

restorer (BLR) circuit is included in this output for

improved performance at all count rates. A switch

on the front panel permits the threshold for the

restorer gate to be determined automatically,

according to the input noise level, or manually, with

a screwdriver adjustment. The switch also has a

center PZ Adj setting that can be used to eliminate

theBLReffectwhenmakingpole-zeroadjustments.
The unipolar output dc level can be adjusted from

-0.1 to +0.1 V to set the zero intercept on the

analyzer when direct coupling is used. The bipolar

output, with a 0 to 10 V range, can be used for

crossover timing or may be preferred for

spectroscopy when operating into ac-coupled

systems at high counting rates. Typical system

block diagrams for a variety of experiments are

described in Section 4.
Three general methods of termination are used.

The simplest of these is shunt termination at the

receiving end of the cable. A second method is

series termination at the sending end. The third is

a combination of series and shunt termination,

where the cable impedance is matched both in

series at the sending end and in shunt at the

receiving end. The combination is most effective,

but this reduces the amount of signal strength at

the receiving end to 50% of that which is available

in the sending instrument.
To use shunt termination at the receiving end of the

cable, connect the <1Ω output of the 572A (on the

front panel) through 93Ω cable to the. input of the

receiving instrument. Then use a BNC tee

connector to attach both the interconnecting cable

and a 100Ω terminator at the input connector of the

Содержание 572A

Страница 1: ...Model 572A Spectroscopy Amplifier Operating and Service Manual Printed in U S A ORTEC Part No 785100 0904 Manual Revision D...

Страница 2: ...nbeassignedtotheunit Also ORTECmust be informed either in writing by telephone 865 482 4411 or by facsimile transmission 865 483 2133 of the nature of the fault of the instrument being returned and of...

Страница 3: ...NS 7 3 7 SHORTING OR OVERLOADING THE AMPLIFIER OUTPUTS 8 3 8 INHIBIT OUTPUT CONNECTION 8 3 9 BUSY OUTPUT CONNECTION 8 3 10 CRM OUTPUT CONNECTION 8 4 OPERATING INSTRUCTIONS 8 4 1 INITIAL TESTING AND OB...

Страница 4: ...a hazard that could result in bodily harm if the safety instruction is not observed CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed Please re...

Страница 5: ...he instrument during external cleaning use only enough liquid to dampen the cloth or applicator SAFETY WARNINGS AND CLEANING INSTRUCTIONS Cleaning Instructions To clean the instrument exterior Unplug...

Страница 6: ...vi...

Страница 7: ...m 100 mV to 100mV Thisoutputpermitstheuse of the direct coupled input of the analyzer with a minimum amount of interface problems The 572A bipolar output may be preferable for spectroscopy when operat...

Страница 8: ...seline with the preamplifier time constant The pole zero adjust is accessible from the front panel of the 572A and can easily be adjusted by observing the baselinewithanoscilloscopewithamonoenergetic...

Страница 9: ...3 Fig 1 1 Differentiation in an Amplifier Without Pole Zero Cancellation Fig 1 2 Differentiation in a Pole Zero Canceled Amplifier...

Страница 10: ...o 5OEC CROSSOVER WALK 3 ns for 50 1 dynamic range includingcontributionofORTEC551or552 Constant FractionTimingSingle ChannelAnalyzer using 50 fraction and 0 5 s shaping COUNT RATE STABILITY The1 33MeV...

Страница 11: ...iometeradjuststheunipolar output baseline dc level range 100 mV to 100 mV 2 3 INPUT INPUT Type BNC front and rear panel connectors accept either positive or negative pulses with rise times in the rang...

Страница 12: ...e pole zero adjustment see Section 4 6 An alternatemethodisaccomplishedeasilybyusinga monoenergetic source and observing the amplifier baseline with an oscilloscope after each pulse under approximatel...

Страница 13: ...h shaping time constant is selected The 572A provides both unipolar and bipolar outputs Theunipolaroutputpulsesshouldbeused in applications where the best signal to noise ratio resolution is most impo...

Страница 14: ...e pileup is sensed in the 572A and the pulse can then be used to prevent the Analyzer from measuring and storing a false amplitude The signal is dc coupled and rises from 0toabout 5Vforatimeequalto6 s...

Страница 15: ...lates the threshold of the BLR gate according to the output noise level The Threshold setting permits manual control of the BLR gate threshold usingthescrewdrivercontrolimmediatelybelowthe toggle swit...

Страница 16: ...on power in the Bin and Power Supply and allow the electronics of the system to warm up and stabilize b Set the 572A controls initially as follows Shaping 2 s Coarse Gain 50 Gain 1 000 Internal Jumper...

Страница 17: ...of the preamplifier Adjusttheattenuatorsothatthe572A output amplitude is about 9 V d Observe the unipolar output of the 572A with an oscilloscope triggered from the 572ABusyoutput Adjust the PZ Adj co...

Страница 18: ...tal deflection Trigger the oscilloscope with the Busy output from the 572A d Reduce the control setting until the baseline discriminator begins to trigger on noise this corresponds to about 200 counts...

Страница 19: ...r for the same pulse height as the pulse obtained in stepc LocktheNormalizecontrolanddonotmove it again until recalibration is required The pulser is now calibrated the Pulse Height dial read directly...

Страница 20: ...capacitance decreases thus reducing the resolution spread The overall resolution spread will depend upon which effect is dominant Figure4 7showscurvesoftypicalnoise resolution spread versus bias volt...

Страница 21: ...PECTROSCOPY SYSTEMS HIGH RESOLUTION ALPHA PARTICLE SPECTROSCOPYSYSTEM Theblockdiagramof a high resolution Spectroscopy system for measuringnaturalalphaparticleradiationisshown in Fig 4 11 Alpha partic...

Страница 22: ...ounters operated at high gas amplification tend to degrade the resolution capabilities drastically at x ray energies even at relatively low counting rates By using a high gain low noise amplifying sys...

Страница 23: ...17 Fig 4 13 Scintillation Counter Gamma Spectroscopy System Fig 4 14 High Resolution X Ray Energy Analysis System Using a Proportional Counter Fig 4 15 General System Arrangement for Gating Control...

Страница 24: ...18 Fig 4 16 Gamma Ray Charged Particle Coincidence Experiment Fig 4 17 Gamma Ray Pair Spectrometry...

Страница 25: ...ant of 0 5 1 2 3 6 or 10 s The amplifier produces the fastshapedpulsefromthesamepreamplifierinput pulse and this triggers discriminator IC12 set just above the noise level The discriminator response t...

Страница 26: ...20 Fig 5 1 Amplifier Block Diagram...

Страница 27: ...ackard 3400A RMS Voltmeter 6 2 PULSER TEST1 Coarse Gain 1K Gain 1 5 Input Polarity Positive Shaping Time Constant 1 s BLR PZ Adj Variable control Fully CW for 300 mV a Connect a positive pulser output...

Страница 28: ...y to vary the PZ Adj control on the front panel in order to cancel the pulser pole and minimize the time required for return to the baseline Increasethepulseroutputamplitudeto1000times the setting tha...

Страница 29: ...rvations 6 4 FACTORY REPAIR This instrument can be returned to the ORTEC factoryforserviceandrepairatanominalcost Our standard procedure for repair ensures the same quality control and checkout that a...

Страница 30: ...served bus 34 Power return ground 13 Spare 35 Reset Scaler 14 Spare 36 Gate 15 Reserved 37 Reset Auxiliary 16 12 V 38 Coaxial 17 12 V 39 Coaxial 18 Spare bus 40 Coaxial 19 Reserved bus 41 117 V ac neu...

Отзывы: