60
Version 4.2: Aug 26, 2015
Turn on the heating system at least 60 minutes before starting the work. If possible place all needed equipment on
the surface during the warming up period to warm these appropriately. Always wait for the temperature to stabilize
completely before starting work.
CAUTION. There is heat loss from the edges of the heated surface. Do not
place temperature sensitive material there.
a. Temperature Controller
The display shows the temperature readout from the chosen channel (only channel 1 is used in this example). In
case of an alarm situation a red light will start flashing in the top left-hand corner of the display. On the right-hand
side of the display it is indicated which parameter is shown in the display. Below is listed all possible values and
messages that can be shown in the display:
Display
Description
OFF
Controller is in standby
OR
Probe T1 out of range or failure
HI
Room high temperature alarm
LO
Room low temperature alarm
TUN
Controller is in auto tuning
E1
In tuning: Timeout 1 error
E2
In tuning : Timeout 2 error
E3
Out of range error
LED
Description
OUT1
Channel 1 output
OUT2
Channel 2 output