![Origio MARS 1200 CLASS II Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/origio/mars-1200-class-ii/mars-1200-class-ii_user-manual_1651530012.webp)
12
Version 4.2: Aug 26, 2015
Some internal components of the workstation may become contaminated during normal use of
the unit. Only experienced personnel competent in decontamination procedures or certified
companies should decontaminate the workstation before servicing these components.
Ensure that the workstation is connected to electrical service in accordance with local and
national electrical codes. Failure to do so may create a fire or electrical hazard.
Do not remove or service any electrical components without first disconnecting the workstation
from electrical service
Avoid the use of flammable gases or solvents in the workstation. Care must be taken to ensure
against the concentration of flammable or explosive gases or vapors.
An open flame should NOT be used in the workstation. Open flames disrupt airflow patterns in
the workstation and may burn the HEPA filter and/or damage the filter’s adhesive.
Gases under high pressure should not be used in the workstation, as they may disrupt its airflow
patterns
HEPA filters only remove particulate matter. Operations generating volatile toxic chemicals or
radionuclides must be evaluated carefully and when approved by an appropriate safety offer may
require connection to a dedicated external exhaust through a thimble connection.
The HEPA filter is fragile and should not be touched. Avoid puncturing either HEPA filter during
installation or normal use. If you suspect that a HEPA filter has been damaged, DO NOT use the
workstation; contact a local certification provider.
Proper operation of the cabinet depends largely upon its location and the operator’s work habits.
Consult the Installation and Operation sections of this manual for further details.
Customer materials which are not UV-resistant should not be used during UV-disinfection inside
the work chamber.
The HEPA filter in the workstation will gradually accumulate airborne particulate matter from the
room and from work performed in the workstation. The rate of accumulation will depend upon the
cleanliness of the room air. The airflow sensors will alarm when the HEPA filters will need to be
checked.
The VOC filter placed in the contained box on top of the workstation (optional) has to be
exchanged every 3 to 6 month. The rate of exchange of the VOC filter will depend upon the
cleanliness of the room air. Only experienced personnel competent in decontamination
procedures or certified companies should replace and decontaminate the workstation before
use.
Fuses must be exchanged by a qualified technician or a certified company. Please contact your
nearest ORIGIO EQUIPMENT distributor for assistance.
Содержание MARS 1200 CLASS II
Страница 1: ...Mars Workstation User Manual...
Страница 2: ...Version 4 2 Aug 26 2015...
Страница 21: ...Version 4 2 Aug 26 2015 21...
Страница 22: ...22 Version 4 2 Aug 26 2015...
Страница 23: ...Version 4 2 Aug 26 2015 23...
Страница 24: ...24 Version 4 2 Aug 26 2015...